1.PCR-based Identification of Eubacteirum species in endodontic infection.
Journal of Korean Academy of Conservative Dentistry 2003;28(3):241-248
No abstract available.
2.Apical prepration size in infected root canals.
Journal of Korean Academy of Conservative Dentistry 2010;35(1):1-4
The final preparation (MAF) size in infected root canals is still controversial. Nonetheless, recent studies demonstrated that larger apical preparation sizes produces a greater reduction in remaining bacteria and dentinal debris as compared to smaller apical preparation sizes. Therefore, clinicians should be practiced with treatment strategies guided by evidence-based information, especially in infected/failed root canals.
Bacteria
;
Dental Pulp Cavity
;
Dentin
;
Tooth Apex
3.Comparative study on morphology of cross-section and cyclic fatigue test with different rotary NiTi files and handling methods.
Jae Gwan KIM ; Kee Yeon KUM ; Eui Seong KIM
Journal of Korean Academy of Conservative Dentistry 2006;31(2):96-102
There are various factors affecting the fracture of NiTi rotary files. This study was performed to evaluate the effect of cross sectional area, pecking motion and pecking distance on the cyclic fatigue fracture of different NiTi files. Five different NiTi files-Profile(R)(Maillefer, Ballaigue, Switzerland), ProTaper(TM) (Maillefer, Ballaigue, Switzerland), K3(R) (SybronEndo, Orange, CA), Hero 642(R) (Micro-mega, Besancon, France), Hero Shaper(R)(Micro-mega, Besancon, France)-were used. Each file was embedded in temporary resin, sectioned horizontally and observed with scanning electron microscope. The ratio of cross-sectional area to the circumscribed circle was calculated. Special device was fabricated to simulate the cyclic fatigue fracture of NiTi file in the curved canal,. On this device, NiTi files were rotated (300rpm) with different pecking distances (3 mm or 6 mm and with different motions (static motion or dynamic pecking motion). Time until fracture occurs was measured. The results demonstrated that cross-sectional area didn't have any effect on the time of file fracture. Among the files, Profile(R) took the longest time to be fractured. Between the pecking motions, dynamic motion took the longer time to be fractured than static motion. There was no significant difference between the pecking distances with dynamic motion, however with static motion, the longer time was taken at 3mm distance. In this study, we could suggest that dynamic pecking motion would lengthen the time for NiTi file to be fractured from cyclic fatigue.
Citrus sinensis
;
Fatigue*
;
Fractures, Stress
4.Effect of surface defects and cross-sectional configuration on the fatigue fracture of NiTi rotary files under cyclic loading.
Yu Mi SHIN ; Eui Sung KIM ; Kwang Man KIM ; Kee Yeon KUM
Journal of Korean Academy of Conservative Dentistry 2004;29(3):267-272
The purpose of this in vitro study was to evaluate the effect of surface defects and cross-sectional configuration of NiTi rotary files on the fatigue life under cyclic loading. Three NiTi rotary files (K3(TM), ProFile(R), and HERO 642(R)) with #30/.04 taper were evaluated. Each rotary file was divided into 2 subgroups: control (no surface defects) and experimental group (artificial surface defects). A total of six groups of each 10 were tested. The NiTi rotary files were rotated at 300rpm using the apparatus which simulated curved canal (40 degree of curvature) until they fracture. The number of cycles to fracture was calculated and the fractured surfaces were observed with a scanning electron microscope. The data were analyzed statistically. The results showed that experimental groups with surface defects had lower number of cycles to fracture than control group but there was only a statistical significance between control and experimental group in the K3(TM) (p<0.05). There was no strong correlation between the cross-sectional configuration area and fracture resistance under experimental conditions. Several of fractured files demonstrated characteristic patterns of brittle fracture consistent with the propagation of pre-existing cracks. This data indicate that surface defects of NiTi rotary files may significantly decrease fatigue life and it may be one possible factor for early fracture of NiTi rotary files in clinical practice.
Fatigue*
;
Fractures, Stress*
5.Scanning electron microscopic study on the efficacy of root canal wall debridement of rotary Ni-Ti instruments with different cutting angle.
In Soo JEON ; Kee Yeon KUM ; Seong Ho PARK ; Tai Cheol YOON
Journal of Korean Academy of Conservative Dentistry 2002;27(6):577-586
The purpose of this in vitro study was to compare the effects of root canal debridement following rotary Ni-Ti instruments with positive versus negative rake angle. Seventy sound, extracted human anterior teeth & premolars were randomly divided into four groups. The used rotary instruments were Ni-Ti HERO 642(Micro-Mega in France, 20 specimen), Ni-Ti ProFile(Maillefer, Ballaigues, Switzerland, 20 specimen), stainless steel engine reamer(Mani, Matsutani Seisakusho Co.,Japan, 20 specimen) and negative control group(10 specimen) was only extirpated with barbed broach(Mani, Matsutani Seisakusho Co.,Japan) Group 1 & 2 teeth were prepared to a #40 at the apex followed by 1 mm using crown-down technique. Group 3 teeth were instrumented from a #15 to a #40 in sequential order. After preparation and final irrigation, the roots split longitudinally into a bucco-lingual direction. Root halves were cross-sectioned in apical third portion again. all root specimens were prepared for SEM investigation & photographed. Separate evaluations were undertaken for smear layer on prepared walls with a five score-index for each using reference photograph in root halves. the penetration depth of smear layer into dentinal tubules was also estimated in the other halves. the following results were obtained : 1. Smear layer was observed on all the prepared walls with three experimental groups except negative control group 2. Smear layer characteristics 1) HERO 642 groups showed snowy & dusty appearance & were observed only few some dentinal tubuli open on the prepared walls, and the penetration depth of it into dentinal tubules may be 1-2 microm thick. 2) ProFile groups showed shiny & burnished appearance & complete root canal wall covered by a homogenous smear layer with no open dentinal tubuli and penetration depth of it into dentinal tubules may be 1-2 microm thick. 3) Engine reamer groups showed obviously file's passed tracks on the prepared walls & were observed complete root canal wall covered by a homogenous smear layer with no open dentinal tubuli. The results revealed that a completely clean root canal could not be achieved regardless of positive & negative rake angle, which is in accordance with the majority of studies on root canal cleanliness. In conclusion, throughout irrigation with antibacterial solutions or chelating agents is recommended to remove the smear layer on prepared canal walls.
Bicuspid
;
Chelating Agents
;
Debridement
;
Dental Pulp Cavity
;
Dentin
;
Electrons
;
France
;
Humans
;
Nickel
;
Smear Layer
;
Stainless Steel
;
Switzerland
;
Titanium
;
Tooth
;
Track and Field
6.Analysis of temperature changes and sterilization effect of diode laser for the treatment of peri-implantitis by wavelength and irradiation time.
Jeong Hwan SEOL ; Jun Jae LEE ; Kee Yeon KUM ; Jong Ho LEE ; Young Joon LIM
Journal of Dental Rehabilitation and Applied Science 2017;33(3):178-188
PURPOSE: We compared the effects of newly developed diode laser (Bison 808 nm Diode laser) on the treatment of peri-implantitis with conventional products (Picasso 810 nm Diode laser) by comparing the surface temperature of titanium disc and bacterial sterilization according to laser power. MATERIALS AND METHODS: The titanium disc was irradiated for 60 seconds and 1 – 2.5 W using diode laser 808 nm and 810 nm. The surface temperature of the titanium disc was measured using a temperature measurement module and a temperature measurement program. In addition, in order to investigate the sterilizing effect according to the laser power, 808 nm laser was irradiated after application of bacteria to sandblasted large-grit acid-etched (SLA) and resorbable blast media (RBM) coated titanium discs. The irradiated disks were examined with scanning electron microscopy. RESULTS: Both 808 nm and 810 nm lasers increased disk surface temperature as the power increased. When the 810 nm was irradiated under all conditions, the initial temperature rise rate, the descending rate, and the temperature change before and after was higher than that of 808 nm. Disk surface changes were not observed on both lasers at all conditions. Bacteria were irradiated with 808 nm, and the bactericidal effect was increased as the power increased. CONCLUSION: When applying these diode lasers to the treatment of peri-implantitis, 808 nm which have a bactericidal effect with less temperature fluctuation in the same power conditions would be considered safer. However, in order to apply a laser treatment in the dental clinical field, various safety and reliability should be secured.
Bacteria
;
Lasers, Semiconductor*
;
Microscopy, Electron, Scanning
;
Peri-Implantitis*
;
Sterilization*
;
Titanium
7.Assessment of sterilization effect and the alteration of surface texture and physical properties of gutta-percha cone after short-term chemical disinfection.
Nan Sim PANG ; Il Young JUNG ; Yoon Jung YU ; Kee Yeon KUM
Journal of Korean Academy of Conservative Dentistry 2006;31(2):133-140
The purposes of this study were firstly to identify the microbial species on gutta-percha (GP) cones exposed at clinics using polymerase chain reaction, and secondly to evaluate the short-term sterilization effect of three chemical disinfectants. It also evaluated the alteration of surface texture and physical properties of GP cones after 5-min soaking into three chemical disinfectants. 150 GP cones from two endodontic departments were randomly selected for microbial detection using PCR assay with universal primer. After inoculation on the sterilized GP cones with the same microorganism identified by PCR assay, they were soaked in three chemical disinfectants: 5% NaOCl, 2% Chlorhexidine, and ChloraPrep for 1, 5, 10, and 30 minutes. The sterilization effect was evaluated by turbidity and subculture. The change of surface textures using a scanning electron microscope and the tensile strength and elongation rate of the GP cones were measured using an Instron 5500 (Canton). Statistical analysis was performed. Four bacterial species were detected in 29 GP cones (19.4%), and all the species belonged to the genus Staphylococcus. All chemical disinfectants were effective in sterilization with just 1 minute soaking. On the SEM picture of NaOCl-soaked GP cone, a cluster of cuboidal crystals was seen on the cone surface. The tensile strength of NaOCl-soaked group was significantly higher than the other groups (p < 0.05). Also, all disinfectants significantly increased the elongation rate of GP cones compared to the fresh GP cone (p < 0.05). Present data demonstrate that three chemical disinfectants are useful for rapid sterilization of GP cone just before obturation.
Chlorhexidine
;
Disinfectants
;
Disinfection*
;
Gutta-Percha*
;
Polymerase Chain Reaction
;
Staphylococcus
;
Sterilization*
;
Tensile Strength
8.The sustaining effect of three polymers on the release of chlorhexidine from a controlled release drug device for root canal disinfection.
Young Bin BOK ; Doug Youn LEE ; Chang Young LEE ; Kyung Nam KIM ; Kee Yeon KUM
Journal of Korean Academy of Conservative Dentistry 2004;29(6):548-554
The aim of this in vitro study was to evaluate the suitability of using chitosan, poly (lactide-co-glycolide) (PLGA), and polymethyl methacrylate (PMMA) to control the release of chlorhexidine digluconate (CHX) from a prototype of controlled release drug device (CRD) for root canal disinfection. Four different prototypes with different formulations were prepared. Group A (n = 12); The device (absorbent paper point) was loaded with CHX as control. Group B (n = 12); same as group A, but the device was coated with chitosan. In Groups C and D, the device was treated in the same way as group A and then coated three times with 5% PMMA (Group C, n = 12), or coated three times with 3% PLGA (Group D, n = 12). The devices were randomly allocated to experimental groups of 12 each. All CRD prototypes were soaked in 3 mL distilled water. The concentrations of CHX were determined using a UV spectrophotometer. The surface characteristics of each prototype were observed using a scanning electron microscope. The result showed that release rate of CHX was the greatest in the non-coated group, followed by the chitosan-coated group, the PLGA-coated group, and the PMMA-coated group (P < 0.05). Pores were observed on the surface of the prototypes that were coated with PLGA and PMMA. When the pore size was smaller, the release rate was lower. This data indicate that polymer coating can control the release rate of CHX from the CRD prototypes.
Chitosan
;
Chlorhexidine*
;
Dental Pulp Cavity*
;
Disinfection*
;
Polymers*
;
Polymethyl Methacrylate
;
Water
9.A retrospective study of the intentionally replanted mandibular second molars with C-shaped root canal configurations.
Won Jun SHON ; Kee Yeon KUM ; Seung Ho BAEK ; Woo Cheol LEE
Journal of Korean Academy of Conservative Dentistry 2011;36(1):19-25
OBJECTIVES: The purpose of this retrospective study was to evaluate the success rate of intentionally replanted mandibular second molar with C-shaped canal configurations and to access the impact of preoperative periapical lesion on the success of intentional replantation procedure. MATERIALS AND METHODS: This retrospective chart review study evaluated 52 intentionally replanted mandibular second molar teeth treated at Seoul National University Dental Hospital Department of Conservative Dentistry from January 2005 to December 2007. Seventeen teeth were lost for the follow-up, and another 6 teeth did not meet inclusion criteria of C-shaped root canal configurations. Healing outcome such as success, uncertain healing, and failure after follow-up was evaluated by clinical criteria and radiographs. RESULTS: The overall success rate was 72.4% for the 29 intentionally replanted C-shaped mandibular second molars. The success rate of replanted teeth with preoperative periapical lesions was similar to that of replanted teeth which have no periapical lesions. CONCLUSIONS: Therefore, root canal treatment failure on C-shaped mandibular second molar can be predictably treated by intentional replantation regardless of the presence of periapical lesion.
Dental Pulp Cavity
;
Dentistry
;
Follow-Up Studies
;
Hospital Departments
;
Intention
;
Molar
;
Replantation
;
Retrospective Studies
;
Tooth
;
Treatment Failure
10.The Effect of Surface Defects on the Cyclic Fatigue Fracture of HEROShaper Ni-Ti rotary files in a Dynamic Model: A Fractographic Analysis.
Jung Kyu LEE ; Eui Sung KIM ; Myoung Whai KANG ; Kee Yeon KUM
Journal of Korean Academy of Conservative Dentistry 2007;32(2):130-137
This in vitro study examined the effect of surface defects on cutting blades on the extent of the cyclic fatigue fracture of HEROShaper Ni-Ti rotary files using fractographic analysis of the fractured surfaces. A total of 45 HEROShaper (MicroMega) Ni-Ti rotary files with a #30/.04 taper were divided into three groups of 15 each. Group 1 contained new HEROShapers without any surface defects. Group 2 contained HEROShapers with manufacturing defects such as metal rollover and machining marks. Group 3 contained HEROShapers that had been clinically used for the canal preparation of 4-6 molars. A fatigue-testing device was designed to allow cyclic tension and compressive stress on the tip of the instrument whilst maintaining similar conditions to those experienced in a clinic. The level of fatigue fracture time was measured using a computer connected the system. Statistical analysis was performed using a Tukey's test. Scanning electron microscopy (SEM) was used for fractographic analysis of the fractured surfaces. The fatigue fracture time between groups 1 and 2, and between groups 1 and 3 was significantly different (p < 0.05) but there was no significant difference between groups 2 and 3 (p > 0.05). A low magnification SEM views show brittle fracture as the main initial failure mode. At higher magnification, the brittle fracture region showed clusters of fatigue striations and a large number of secondary cracks. These fractures typically led to a central region of catastrophic ductile failure. Qualitatively, the ductile fracture region was characterized by the formation of microvoids and dimpling. The fractured surfaces of the HEROShapers in groups 2 and 3 were always associated with pre-existing surface defects. Typically, the fractured surface in the brittle fracture region showed evidence of cleavage (transgranular) facets across the grains, as well as intergranular facets along the grain boundaries. These results show that surface defects on cutting blades of Ni-Ti rotary files might be the preferred sites for the origin of fatigue fracture under experimental conditions. Furthermore, this work demonstrates the utility of fractography in evaluating the failure of Ni-Ti rotary files.
Edible Grain
;
Fatigue*
;
Fractures, Stress*
;
Microscopy, Electron, Scanning
;
Molar