Failure analysis was performed to investigate the failure of the femur fixation plate which was previously fixed on the femur of a girl. Radiography, metallography, fractography and mechanical testing were conducted in this study. The results show that the failure was due to the formation of notches on the femur plate. These notches act as stress raisers from where the cracks start to propagate. Finally fracture occurred on the femur plate and subsequently, the plate failed.
*Bone Plates
;
Bone Screws
;
*Equipment Failure Analysis
;
Femur/radiography
;
Femur/*surgery
;
Fracture Fixation, Internal/*instrumentation
;
*Stainless Steel
;
Surface Properties