1.MRI Imaging Device Channel Number Acceptance Appraisal Research
Chinese Journal of Medical Instrumentation 2016;40(3):217-220
This paper verifies MRI RF system conformity of the actual channel number and the manufacturer's nominal channel number, and inspects the actual channel number’s effect on the properties of magnetic resonance imaging (MRI). At the same time through detecting the increasing channel number’s relationship with MRI detection system image noise level and the image reconstruction speed, to verify the number of MRI RF system receiver is closely related with SNR, MRI scan time and so on.
2.Research on the Evaluation System of Clinical Value of Digital Radiography.
Fan BI ; Bin LI ; Houde CAO ; Haisheng HU ; Wenyong TU
Chinese Journal of Medical Instrumentation 2019;43(2):146-149
OBJECTIVE:
To establish a set of evaluation system of digital radiography clinical value and provide foundation for the maturity assessment of digital radiography.
METHODS:
The evaluation system of clinical value of digital radiography was established by literature survey,expert consultation,and percentage weight method.
RESULTS:
The expert authority coefficients were 0.81 and 0.88,respectively.After two rounds of consultation,variation coefficients of each item ranged from 0 to 0.207,and the coefficient coordination were 0.599.The index system consisted of 5 first-level indexes and 12 second-level indexes.The weights of first-level indexes such as image quality,safety,usability,economic and social indicators share of the weight are 0.298,0.294,0.199,0.121 and 0.088 respectively.
CONCLUSIONS
A completed and scientific evaluation system was established,which provides a scientific assessment tool for clinical value of digital radiography.
Delphi Technique
;
Radiographic Image Enhancement
;
standards