1.Delayed migration of K-wire into popliteal fossa used for tension band wiring of patellar fracture.
Sanjay MEENA ; Hira-Lal NAG ; Senthil KUMAR ; Nilesh BARWAR ; Samarth MITTAL ; Amit SINGLA
Chinese Journal of Traumatology 2013;16(3):186-188
Breakage of K-wires and stainless steel wires which are used for fracture fixation is not uncommon, but migration is rare. We report a case of migration of broken K-wire used for patella tension band wiring to the popliteal fossa. The broken hardware was removed surgically. We would like to suggest that K-wire and wire fixation used for treatment of patellar fractures can migrate into the posterior compartment of the knee and cause clinical symptoms. Close clinical and radiological follow-up after internal fixation to identify the presence of hardware breakage or movement and removal of wires once fracture has united can avert such complications.
Adult
;
Bone Wires
;
Equipment Failure
;
Foreign-Body Migration
;
etiology
;
Fracture Fixation, Internal
;
instrumentation
;
methods
;
Fractures, Bone
;
diagnostic imaging
;
surgery
;
Humans
;
Male
;
Patella
;
diagnostic imaging
;
injuries
;
Radiography
;
Time Factors