1. Study on the design and application of nasal endoscope screw self-open cleaning loading technology
Fenglian CHE ; Minzhi JING ; Liming QIAN
Chinese Journal of Practical Nursing 2019;35(18):1410-1412
Objective:
To improve the cleaning and loading technology of nasal endoscope in order to ensure the cleaning effect, to improve the working efficiency, and to reduce the cleaning cost.
Methods:
A total of 480 reusable contaminated nasal endoscopes were divided into two groups according to random number table: Clean and load the nasal endoscopes (group A) with the usage of metal parts to open up the leaf; Clean and load the nasal endoscope (group B) with the usage of screw driver to open up the leaf. Each group has 240 pieces. After mechanical cleaning, observe and analyze the cleaning effect, loading effectiveness, loading capacity, loading time and CSSD operators′ degree of satisfaction.
Results:
The number of qualified cleaning joint parts of the two groups were 218 and 231 respectively, and the number of qualified cleaning leaf of the two groups were 214 and 227 respectively, and there were significant differences between the two groups (
2.Application effect of silica gel stent loading technology in precision microscopic instrument management
Fenglian CHE ; Weiwei CHEN ; Liming QIAN
Chinese Journal of Practical Nursing 2022;38(26):2028-2032
Objective:To improve the loading technology of precision micro instruments, so as to ensure not only the quality of cleaning and sterilization, but also the safe use of instruments.Methods:From January to March, 2021, 80 sets of precision micro instruments from Sterilization Supply Center, Shanghai Jiao Tong University School of Medicine Affiliated Ruijin Hospital were randomly divided into silica gel pad loading group (group A) and silica gel stent loading group (group B), 40 sets in each group. The cleaning effect, drying effect, loading effectiveness and sterilization wet bag of the instruments in the two groups were compared, and the staff of the disinfection supply center were satisfied with the operation of the loading technology of the instruments in the two groups.Results:The qualified cleaning numbers of the instruments in group A and B were 37 and 39 respectively, and there was no significant difference between the two groups ( χ2 = 1.05, P>0.05). The qualified drying numbers in the group A and B were 29 and 38 respectively, the difference between the two groups was significant ( χ2 = 7.44, P<0.05). The effective number of loading in the group A and B were 30 cases and 39 cases respectively, the difference was significant between the two groups ( χ2 = 8.54, P<0.05). The number of sterilized wet bags in the group A and B were 6 cases and 0 cases respectively, the difference was significant between the two groups ( χ2 = 6.49, P<0.05). The satisfaction of 44 personnel in the disinfection supply center with the operation of loading technology were 36 and 43 respectively, the difference was statistically significant between the two groups ( χ2 = 6.07, P<0.05). Conclusion:Loading precision micro instruments with silica gel stent can not only ensure the cleaning and sterilization quality of instruments, but also make the loading of instruments more safe and stable, greatly improve the working efficiency of instrument treatment and accelerate the turnover efficiency of instruments.
3.Comparative effectiveness research of different testing methods in functional quality inspection of precision surgical instruments
Fenglian CHE ; Liming QIAN ; Weiwei CHEN ; Lele ZHENG ; Yu CHEN ; Huiyuan PAN
China Medical Equipment 2024;21(3):129-132
Objective:To compare the testing effects of different testing methods in the functional inspection of precision surgical scissors and needle holders.Methods:By consulting relevant literature and professional books,three testing methods and materials for precision surgical scissors and needle holders were selected.The cutting performance of the precision surgical scissors were tested by gauze cloth,cotton wool,and silicone film method in the Interlock surgical instrument testing set.The clamping performance of precision needle holder was tested by using light,needle clips,and thread clamping method in the Interlock surgical instrument testing set.100 pieces of precision surgical scissors and needle holders which were qualified for cleaning and disinfection in the hospital were selected for effectiveness testing,and a self-made satisfaction questionnaire to investigate the satisfaction of surgeons with the use of instruments.Results:After 100 precision surgical scissors tested using three methods of gauze,cotton wool,and silicone film,the qualified number of cutting performance was 94,76,and 58,respectively,the difference was statistically significant(x2=36.526,P<0.05).The satisfaction rate of surgeons with the use of precision surgical scissors passed the silicone film test was the highest at 100%,followed by cotton wool cloth at 82.0%,and the gauze method at the worst at 67.0%,the difference was statistically significant(x2=48.439,P<0.05).After 100 precision needle holders were tested by using three methods of light irradiation,needle clamping,and wire clamping,the qualified number of clamping performance was 96,78,and 62,respectively,the difference was statistically significant(x2=38.160,P<0.05).The satisfaction rate of surgeons with the precision needle holders passed the clamp line test was the highest at 100.0%,followed by the clamp needle at 79.0%,and the worst under light at 62.0%,the difference was statistically significant(x2=57.705,P<0.05).Conclusion:The function and quality of precision surgical scissors and needle holders tested by the test objects in Interlock Surgical Instrument Test Suite can objectively and accurately reflect the functional status of surgical scissors and needle holders,provide reference for regular maintenance and upkeep of instruments,and ensure safe use of instruments.