1.Comparison of two chemiluminescence systems used in the measurement of plasma aldosterone and renin concentrations
Xiuyong YANG ; Fengfan ZHENG ; Ying SONG ; Shumin YANG ; Jinbo HU ; Wenwen HE ; Zhi-Hong WANG ; Qingfeng CHENG ; Qifu LI
Chinese Journal of Endocrine Surgery 2019;13(5):413-418
Objective To compare the imported (Diasorin,Italy) and domestic (Mindray,Shenzhen) chemiluminescent systems used in the measurement of plasma aldosterone and renin concentrations;To establish the reference interval of plasma aldosterone and renin concentrations in healthy adults.Methods With the assay instrument and its kits from Italy Diasorin as the reference system,the concentrations of plasma aldosterone and renin were measured by the two systems,in 143 healthy adults,72 patients with hypertension (16 patients with primary aldosteronism) and to establish the medical reference range (P2.5-P97.5) of them.Results The plasma aldosterone (r=0.914,P<0.01) and renin(r=0.977,P<0.01)concentrations detected by the two systems were positively correlated.Distribution of plasma aldosterone and renin was skewed in healthy adults.The reference interval was 30.8-344.6 pg/ml for aldosterone and 2.4-90.0 μIU/ml for renin by the imported chemiluminescent system.The reference interval was 29.4-473.3 pg/ml for aldosterone and 3.6-98.3 μIU/ml for renin by the domestic chemiluminescent detection system.Conclusion The two systems are closely correlated in measuring plasma aldosterone and renin concentrations.
2. On the consistency of different methods for detecting aldosterone concentration in blood
Fengfan ZHENG ; Ying SONG ; Shumin YANG ; Jinbo HU ; Wenwen HE ; Zhihong WANG ; Qingfeng CHENG ; Qifu LI
Chinese Journal of Endocrinology and Metabolism 2019;35(11):934-938
Objective:
To evaluate the consistency of different methods for detecting aldosterone concentration in blood and to establish a reference interval of serum aldosterone concentration by liquid chromatography-tandem mass spectrometry(LC-MS/MS).
Methods:
Concentrations of blood aldosterone were measured by LC-MS/MS, chemiluminescent assays (Diasorin, Domestic A and B systems) and radioimmunoassay (RIA) in 138 healthy adults, 67 patients with essential hypertension and 23 patients with primary aldosteronism.
Results:
Aldosterone concentrations measured by various methods were quiet different(