1.Development of a Three-Wavelength Brain Tissue Oxygen Monitoring System Based on Near Infrared Spectrum
Zexi LI ; Hanlin LI ; Qi YIN ; Shijie CAI ; Jilun YE ; Xu ZHANG ; Hui YU ; Dahai GOU
Chinese Journal of Medical Instrumentation 2024;48(1):26-29,37
In the past 20 years,near infrared spectrum technology has been widely used in human body monitoring due to its non-invasive and real-time characteristics.Oxygen,as the main metabolic substance of the human body,is consumed the most in brain tissue.In order to prevent complications caused by a decrease in brain tissue oxygen during treatment,the patient's brain tissue blood oxygen saturation needs to be monitored in real time.Currently,most of the clinically used non-invasive cerebral blood oxygen detection equipments use dual wavelengths.Other substances on the detection path will cause errors in the measurement results.Therefore,this article proposes a three-wavelength method based on the basic principle of non-invasive monitoring of cerebral blood oxygen using near-infrared spectrum.The brain tissue oxygen saturation monitoring method of detecting light sources was initially verified through the built system,laying the foundation for subsequent system engineering.