2.Application of occluders in endovascular repair of aortic aneurysms.
Zhen-yu SHI ; Wei-guo FU ; Yu-qi WANG ; Da-qiao GUO ; Bin CHEN ; Jun-hao JIANG ; Xin XU ; Jue YANG ; Ting ZHU
Chinese Medical Journal 2007;120(15):1366-1368
3.Endovascular stent-graft exclusion of giant and long-segment thoracic aortic aneurysms.
Shen-ming WANG ; Guang-qi CHANG ; Zuo-jun HU ; Chen YAO ; Xiao-xi LI
Chinese Journal of Surgery 2005;43(18):1191-1194
OBJECTIVETo discuss the availability of endovascular stent-graft exclusion in giant and long-segment thoracic aortic aneurysms.
METHODSEndovascular stent-graft exclusion was performed on 3 male patients with giant and long-segment thoracic aortic aneurysm. Multiple stent-grafts were connected as a long stent-graft to repair these thoracic aortic aneurysms, of which one with 28 cm in length and 7.3 cm in maximum diameter was excluded by 4 stent-grafts with different diameters and same length of 130 mm. The preliminary bypass between right common carotid artery and left common carotid artery or left common carotid artery and left subclavian artery was performed on 2 patients.
RESULTSThree patients underwent the operation successfully, and achieved the good results. There was no endoleak at the conjunction of stent-grafts in 2 cases, and a little endoleak in 1 case, which disappeared 1 year after operation. One case presented temporal cerebral ischemia, which disappeared after 1 month.
CONCLUSIONSThe endovascular deployment with multiple stent-grafts connection is a safe, mini-invasive and effective therapeutic method for giant and long-segment thoracic aortic aneurysm, but long-term outcomes will be gotten after the follow-up study.
Aged ; Aortic Aneurysm, Thoracic ; surgery ; Blood Vessel Prosthesis ; Blood Vessel Prosthesis Implantation ; instrumentation ; methods ; Follow-Up Studies ; Humans ; Male ; Middle Aged ; Stents ; Treatment Outcome
4.The endovascular repair of ruptured aortic aneurysm.
Wei GUO ; Lu-yue GAI ; Xiao-ping LIU ; Guo-hua ZHANG ; Fa-qi LIANG ; Rong LI
Chinese Journal of Surgery 2005;43(18):1187-1190
OBJECTIVETo discuss the safety, feasibility, efficacy and problem of endovascular repair (EVR) for ruptured aortic aneurysm (RAA).
METHODSFourteen patients with RAA were obtained diagnosis and anatomic data of proximal and distal neck through magnetic resonance imaging and CT. The patients with hypotension were accepted anti-shock treatment. Stent-grafts were deployed in proper position of RAA under X-ray fluoroscopic.
RESULTSThe time from rupture to operation was 50 min to 21 d. Three cases had plenty of blood in left pleural cavity and 2 cases were done under bleeding shock condition. Five cases with RAA were fixed by bifurcated stent-grafts and others by straight stent-grafts. Follow-up term was 1 month to 38 months. One with ruptured thoracic aortic aneurysm died 4 h later after operation, another one died of left thoracic cavity infection 9 months later, and others lived well without complication.
CONCLUSIONThe EVR is a safe, efficient and feasible method. It will be a potential alternative to treat RAA in future, but more suitable for RAA with better proximal and distal landing zones.
Adult ; Aged ; Aortic Rupture ; diagnosis ; surgery ; Blood Vessel Prosthesis ; Blood Vessel Prosthesis Implantation ; instrumentation ; methods ; Female ; Humans ; Male ; Middle Aged ; Retrospective Studies ; Stents ; Treatment Outcome
5.Endovascular exclusion of juxtarenal abdominal aortic aneurysm with one-piece customized fenestrated endovascular stent-graft.
Zai-ping JING ; Liang-xi YUAN ; Xiang FENG ; Jun-min BAO ; Zhi-qing ZHAO ; Rui FENG ; Zhi-jun MEI ; Ming-fang LIAO ; Yi-fei PEI
Chinese Journal of Surgery 2007;45(23):1596-1599
OBJECTIVETo evaluate the technical feasibility of juxtarenal abdominal aortic aneurysm (AAA) repair with fenestrated endovascular stent-graft
METHODSA 64-year-old male was diagnosed with juxtarenal AAA with severe coronary artery stenosis, fenestrations was customized according to precise helical CT data to accommodate visceral and renal arteries. Under general anesthesia and dynamic supervision of digital subtraction angiography (DSA), juxtarenal AAA was excluded with the customized fenestrated stent-graft and balloon expandable mini stent-grafts were deployed into bilateral renal arteries respectively.
RESULTSAfter operation, DSA showed the patency of the super mesenteric artery, bilateral renal arteries and left hypogastric artery, no endoleak was found. The serum creatinine decreased slightly after operation. CT angiography revealed favorable morphology of the stent-graft without tortuosity, migration, disjoint and endoleak 10 days after the operation and patency of super mesenteric artery, bilateral renal arteries and left hypogastric artery.
CONCLUSIONSThe placement of customized fenestrated endovascular stent-graft is a feasible option for juxtarenal AAA.
Aortic Aneurysm, Abdominal ; surgery ; Blood Vessel Prosthesis Implantation ; instrumentation ; methods ; Feasibility Studies ; Humans ; Male ; Middle Aged ; Stents
7.Acute phase reaction of different macromolecule vascular grafts healing in rat muscle.
Weici WANG ; Bi JIN ; Chenxi OUYANG ; Yiqing LI ; Weilin XU ; Hongjun YANG ; Haiye XU
Chinese Journal of Biotechnology 2010;26(1):79-84
To find out which biomaterial had the best biocompatibility, we compared the acute phase reaction of common biomaterials preparing for vascular grafts with the material of polyurethane modified by silk fibroin (SF-PU(1:1)). After transplanted the materials of dacron, polyterafluoroethylene (e-PTFE), polyurethane (PU), SF-PU(1:1) in rat muscle for one week, we studied the influence of different biomaterials on the histocompatibility by using rat acute toxicity test, test of local reaction in muscle, tissue section staining, WBC and PLT count. As a result, dacron had the worst histocompatibility. The other biomaterials had slight local inflammatory reaction. The WBC and PLT was nearly the same with the blank except dacron. e-PTFE, pure PU and SF-PU(1:1) had the better histocompatibility than traditional dacron. Especially SF-PU(1:1) had the best histocompatibility. Because of the better physical properties and histocompatibility of SF-PU( 1:1), the prospect of preparing small-diameter vascular grafts with SF-PU was cheerful.
Acute-Phase Reaction
;
Animals
;
Biocompatible Materials
;
Blood Vessel Prosthesis
;
Blood Vessel Prosthesis Implantation
;
instrumentation
;
Female
;
Fibroins
;
chemistry
;
Implants, Experimental
;
Macromolecular Substances
;
chemistry
;
Male
;
Materials Testing
;
methods
;
Muscle, Skeletal
;
physiology
;
Polyurethanes
;
chemistry
;
Prosthesis Design
;
Random Allocation
;
Rats
;
Rats, Sprague-Dawley
8.Total percutaneous endovascular abdominal aortic aneurysm repair.
Hong-peng ZHANG ; Wei GUO ; Xiao-ping LIU ; Tai YIN ; Xin JIA ; Jiang XIONG ; Xiao-hui MA
Chinese Journal of Surgery 2010;48(24):1855-1858
OBJECTIVETo determine the safety and efficacy of total percutaneous endovascular abdominal aortic aneurysm repair using the Perclose ProGlide suture-mediated closure system.
METHODSFrom May 2008 to April 2010, 36 abdominal aortic aneurysm patients were undergone total percutaneous endovascular repair. There were 30 male and 6 female patients with a mean age of 68 years. Endografts used included 3 Endurant endografts, 13 Talent endografts, and 20 Zenith endografts. Prior to insertion of the introducer sheath, two ProGlides were pre-set to 18 to 24 F access sites and one to 14 to 16 F access sites. At last, suture the arteriotomy by tying down knots of the ProGlide following removal of the sheath. Technical success, complications, and procedure and access closure times were evaluated. Follow-up protocol consisted of computed tomography angiograms performed at 3, 6, 9, 12 months, and annually thereafter.
RESULTSTwenty patients were operated under local anesthesia and 16 patients under general anesthesia. A total of 68 femoral arteries were closed with 128 devices. Thirty-eight vessels were with 2 devices, while 8 arteries required 3 devices and 2 arteries required 4 devices for hemostasis and an additional 20 vessels only required a single device. Sixty-three (63/68, 92.6%) vessels were closed successfully. Two vessels converted to open closure. Three vessels complicated with hematomas without surgical procedure. The mean follow-up was (12±3) months. There was one asymptomatic femoral artery dissection 3 months after operation.
CONCLUSIONSTotal percutaneous endovascular abdominal aortic aneurysm repair is safe and effective. But it should be performed at hybrid operating room where can convert to open procedure if necessary.
Aged ; Aged, 80 and over ; Aortic Aneurysm, Abdominal ; surgery ; Blood Vessel Prosthesis Implantation ; instrumentation ; methods ; Female ; Follow-Up Studies ; Humans ; Male ; Middle Aged ; Suture Techniques ; instrumentation ; Treatment Outcome
9.Suitability of Endovascular Repair with Current Stent Grafts for Abdominal Aortic Aneurysm in Korean Patients.
Kay Hyun PARK ; Cheong LIM ; Jae Hang LEE ; Jae Suk YOO
Journal of Korean Medical Science 2011;26(8):1047-1051
Suitability rate of endovascular aneurysm repair (EVAR) and the anatomic features causing unsuitability have not been well determined in Asian patients who have abdominal aortic aneurysm (AAA). In a single Korean center, a total of 191 patients with abdominal aortic aneurysm (maximal diameter > or = 4 cm) were identified. Aortoiliac morphologic characteristics in contrast-enhanced computed tomography images were retrospectively reviewed to determine suitability for EVAR with four FDA-approved stent-grafts. AAA was considered ideally suitable for EVAR in 46.6% of patients. The most frequent causes for unsuitability were common iliac artery (CIA) aneurysm (61.8%) and excessive neck angulation (52.9%). Problems such as small and/or short neck and small access were found in minor incidences. If CIA aneurysm is dealt by overstenting with sacrifice of internal iliac artery, suitability rate can increase to 65%. Larger aneurysms were more frequently unsuitable for EVAR and had more chance of having multiple unfavorable features. In conclusion, the overall feasibility rate for EVAR in Korean patients was not different from that in Western patients. However, considering the difference in the major causes of unsuitability, more attention has to be paid to neck angulation and CIA aneurysm to provide EVAR for more Korean patients especially who have large aneurysm.
Aged
;
Aged, 80 and over
;
Aortic Aneurysm, Abdominal/*surgery
;
*Blood Vessel Prosthesis
;
Blood Vessel Prosthesis Implantation/*instrumentation
;
Female
;
Humans
;
Iliac Aneurysm/surgery
;
Iliac Artery
;
Male
;
Middle Aged
;
Republic of Korea
;
Retrospective Studies
;
*Stents
;
Tomography, Spiral Computed
10.Application of endovascular thoracic branched aortic stent-grafts in the treatment of aortic arch dissection.
Chao LI ; Yu-liang LI ; Zhong-gao WANG ; Qiang ZHANG ; Yong-quan GU ; Jie-fang BIAN
Chinese Journal of Surgery 2005;43(18):1184-1186
OBJECTIVETo report the initial clinical experience of endovascular thoracic branched stent grafts in the treatment of aortic arch dissections involving the left subclavian artery.
METHODSFrom February 2004 to June 2004, 14 patients were cured with the endovascular thoracic branched aortic stent-grafts made by Beijing YuHengJia SciTech Co. All patients had Stanford type B aortic dissection with the entry tears just beyond the origin of the left subclavian artery by an average distance of 8.7 mm. The branched stents were consisted of the aortic section and the branched section. The diameter of the stents was 15% to 20% larger than the diameter of the landing zones of native arteries. The repair procedure was performed in angiography laboratory. The branched stent grafts were delivered under fluoroscopic guidance and implanted into the aortic arch including the left subclavian artery.
RESULTSFourteen branched stent-grafts and 2 additional flexible stent-grafts were delivered successfully in all 14 cases. The entry tears were excluded completely, and the truth lumen of the dissection was revealed to the normal diameter in all patients. Neither peripheral complication nor death occurred. All 14 patients had recover the normal life.
CONCLUSIONIt demonstrates that it is possible to apply the technical feasibility of endovascular thoracic branched aortic stent graft to repair the intimal tear of dissection just beyond the left subclavian artery.
Adult ; Aged ; Aneurysm, Dissecting ; surgery ; Aortic Aneurysm, Thoracic ; surgery ; Blood Vessel Prosthesis ; Blood Vessel Prosthesis Implantation ; instrumentation ; methods ; Female ; Follow-Up Studies ; Humans ; Male ; Middle Aged ; Stents ; Treatment Outcome