OBJECTIVE:
Screen seventeen benzodizepines and their metabolitesin urine by GC/ECD and GC/MS.
METHODS:
They were GC (GC/ECD, GC/MS) assay of benzodizepines and GC (GC/ECD, GC/MS) assay of benzophenones of acid-hydrolytic products of 1,4-benzodizepines.
RESULTS:
The methods were simple and sensitive. The recoverys were 60% to 90% of most benzodizepines, linear calibration curves were 20 ng/ml-200 ng/ml (r > 0.99), and detection limits were 0.5 ng/ml-10 ng/ml.
CONCLUSION
They methods were evaluated with human urine samples.
Anti-Anxiety Agents/urine*
;
Benzodiazepines/urine*
;
Benzophenones/urine*
;
Chromatography, Gas
;
Gas Chromatography-Mass Spectrometry
;
Humans