1.Effects of polylactic acid-polyglycolic acid-polyethylene stent on matrix metalloproteinase 2,9 in rabbits with atherosclerosis
Journal of Jilin University(Medicine Edition) 2006;0(02):-
0.05).However the MMP-2 and pro MMP-2 concentrations were still higher 30 d after implantation(13 935?2 167,15 628?1 739 INT?mm~2 vs 11 568?2 219,13 649?1 894 INT?mm~2,P
2.Comparison of the energy response for several dosimeters used in mammography
Jianchao WANG ; Baorong YUE ; Kedao WEI
Chinese Journal of Radiological Medicine and Protection 2008;28(2):188-190
Objective To evaluate whether eight kinds of dosimeters satisfy the clinical mammography dose measurement through the comparison of energy response.Methods According to GB/T 19629-2005,the energy responses of eight available mammography dosimeters were determined for ten standard radiation qualities in the range 25-40 kVp at the SSDL using a refrence-class dosimeter.Results Except 2000S,the energy response factors of all the other dosimeters were consistent within ±5%.Conclusions The energy response of ionizing chamber type is better than other types of dosimeters.
3.Evaluation performance of calibration apparatus for dosimeters used in X-ray mammography
Jianchao WANG ; Baorong YUE ; Kedao WEI
Chinese Journal of Radiological Medicine and Protection 2015;35(5):371-375
Objective To fulfill the requirements for uncertainty of the calibration apparatus for dosimeters used in X-ray mammography through setting standard radiation quality at the SSDL and developing calibration procedures.Methods According to IEC 61267-2005 and IAEA TRS No 457 to recommend RQR-M and RQA-M series standard radiation quality,the calibration apparatus was evaluated for long term stability of the radiation field over 8 years from 2006 to 2014,including 10 response quantities,such as field homogeneity,change rate of mean air kerma and scatter radiation contributions and so on.In addition,the reference dose instrument was traced back to the PSDL of PTB in Germany by post during 2008 and 2012.Results The field homogeneity (φ 40 mm) relative error was ± 1.4%.The long term stability of the calibration apparatus was less than ± 2% (limits of variation).The scatter radiation contributions at their points of test were below 0.12%.The calibration factors traced to PTB were 0.999-1.000.As a result of the calibration apparatus,the expanded uncertainty was ± 3% (k =2,95% confidence interval).Conclusions The calibration apparatus may meet the requirements of IEC 61267-2005 and IAEA TRS No 457 and has obtained the license of metrology from national regulatory authority.The laboratory now performs very well to calibrate dosimeters used in X-ray mammography.
4.Evaluation and test of 192Ir air kerma strength for afterloading systems
Suming LUO ; Zhijian HE ; Lei GAO ; Baorong YUE
Chinese Journal of Radiological Medicine and Protection 2011;31(1):76-78
Objective To study the method of measuring air kerma strength of afterloading units with 192Ir source by using well type ionization chamber.MethodsThe air kerma strength of 30 afterloading units with 192Ir source was measured using 2000A electrometer and 1000 plus well type ionization chamber,and apparent activity of the source was calculated with the air kerma strength and apparent activity conversion factor.The measured activity of the source was compared with the original value of the source provided by the manufacturer,and the relevant deviation should be within ± 5%.Results Theair kerma strength of afterloding units with 192Ir sources was tested.The relevant deviation of the measured activity and the original value was within -0.1%-4.4%.Conclusions The measurement method with a well type ionization chamber is convenient and highly accurate which can be used for the test of quality control in hospitals.
5.Energy dependence of three passive solid detectors on low energy X-ray
Jianchao WANG ; Hui XU ; Kedao WEI ; Baorong YUE
Chinese Journal of Radiological Medicine and Protection 2010;30(6):735-736
Objective To assess the feasibility of ESD measurements in the mammography by the comparative study of energy dependence of the three passive solid detectors on low energy X-ray.Methods Two thermoluminescent TL) detectors ( LiF(Mg,Ti) and LiF( Mg,Cu, P) and one optically stimulated luminescence (OSL) detector( Al2O3: C ) were irradiated for the same air kerma by ten kinds of beam quality between 25-40 kV in a Mo/Mo low energy X-ray generator,respectively. Results The energy dependences of TL and OSL were 25% and 11% between 25-40 kV, respectively. Conclusions considerable potential for OSL detector can be used in routine quality control and ESD measurements in the mammography.
6.Quality control testing and evaluation for digital mammography
Hui XU ; Ran WAN ; Li GUO ; Jianchao WANG ; Baorong YUE
Chinese Journal of Radiological Medicine and Protection 2010;30(1):89-92
Objective To carry out the quality control testing and evaluation for three digital mammography systems.Methods The performance of three digital mammography systems was assessed by applying methods recommended in the European guidelines for quality assurance in breast cancer screening and diagnosis and Chinese specification for testing of quality control in X-ray mammography.The performance of X-ray generator of three digital mammography systems were tested and evaluated.CDMAM 3.4 phantom with four different thickness(30,40,50,60 mm) were exposured in DR,PCM,and CR system,respectively.The average glandular dose (AGD) value was measured and image quality figure (IQF) analysis was performed in each thickness.Results The X-ray machine performance of DR and CR was in accordance with existing standard,however the standard was inappropriate to evaluate part of X-ray machine performance of PCM system.The AGDs for system DR were 1.20,1.42,1.75 and 2.20 mGy for 30,40,50 and 60 mm PMMA thickness,respectively.The respective AGDs for system PCM and CR were 0.82,1.19,1.33,1.70 mGy and 0.59,0.88,1.47,2.19 mGy.For the same phantom thickness sequence,the IQFs were 21.36,21.57,27.25 and 30.58 for system DR,28.02,29.10,35.90,and 41.24 for system PCM,whereas they were 39.78,39.30,43.85 and 48.08 for system CR.Conclusions The AGDs of all three systems were in accordance with the values recommended in European guideline.The AGD and IQF could provide an effective way for performance assessment and constancy checks for digital mammography systems.
7.Radioactivity analyses of food and drinking water in China following the Fukushima nuclear accident
Yanqin JI ; Liangliang YIN ; Qing TIAN ; Baorong YUE ; Xu SU
Chinese Journal of Radiological Medicine and Protection 2012;32(2):125-128
Objective To summarize the analytical results of radioactivity in the food and drinking water nationwide following the Fukushima nuclear accident,and to evaluate its possible contamination to the public health in China.Methods According to the national standard methods and IAEA,FDA correlative references,the scheme was established on sampling and measurements in food and drinking water after the breakout of the accident.The quality control was requested on the sampling,analyses and data report.Results Trace artificial radioactive isotope of 131I was measured in spinach samples on 2 April 2011 in Beijing. Subsequently 131I was found in 10 kinds of growing leaves vegetables (open field)nationwide.The maximum detectable activity of 131I in vegetables was about 3.1 Bq/kg.Since 3 May 2011,the concentration of 131I has been below the detection limits.No artificial radionulide was detectable in all of milk,drinking water and marine products samples during March to December,2011.Conclusions The food and drinking water measurements in China following the Fukushima nuclear accident denoted that the minor amounts of 131I in vegetables might result in very low absorbed dose and induce no impact on human health.The maximum detectable activity of 131I in vegetables was close to that reported in European countries,and much less than that measured in China immediately after the Chernobyl accident in 1986.
8.Comparison of dosimetric characteristics of two radiochromic films (XR-RV3 and EDR2)
Hui XU ; Jianchao WANG ; Kedao WEI ; Baorong YUE
Chinese Journal of Radiological Medicine and Protection 2014;34(2):145-148
Objective To study the dosimetric characteristics of two different radiochromic films used to estimate peak skin dose(PSD) of patients.The characteristics of these two films were investigated and compared after exposure to ionizing radiation in the diagnostic energy range,including post-exposure gray growth,sensitivity,energy dependence and dose response.Methods GafChromic XR-RV3 film and KODAK EDR2 film were exposed to air kerma 800 mGy free-in-air using five X-ray beam qualities (60,80,100,120 and 140 kVp) in a SSDL.The measurement for each energy was normalized to 80 kV to analyze energy dependence of films.The films were calibrated to different dose level (0.025-10 Gy) onphantom by 80 kV X-rays.The response curve were plotted to analyze sensitivity and dose response.The films were scanned with Epson V750 commercial flatbed scanner.Color channel analysis was performed.Results The post-exposure gray growth of XR-RV3 film was found to be fairly stable.The change were 2%,4%,6% at 24 h,72 h and 6 weeks after exposure respectively.EDR2 film was found to be more sensitivity than XR-RV3 film in low dose.The energy response of the XR-RV3 film and EDR2 film were within 9% and 23% over the clinical diagnostic x-ray energies,respectively.In the dose range of 0.025-10 Gy,for the XR-RV3 film,the red channel with the dose response curve was most obvious.For EDR2 film,the pixel value of three color channels was coincident.The EDR2 film appeared to be saturated when receiving doses greater than 500 mGy.Conclusions The XR-RV3 film is superior to EDR2 film in gray growth,energy dependence,dose-response and other aspects.This film is very suitable for measuring and analyzing PSD of patients in interventional radiology procedures.
9.Development of measurement method using TLD for workers occupation personally exposed to 125Ⅰ seed source in the implant
Suming LUO ; Zhijian HE ; Jilong YUAN ; Baorong YUE ; Kedao WEI
Chinese Journal of Radiological Medicine and Protection 2011;31(4):398-403
Objective To explore the method for measuring and calculating both absorbed dose and effective dose received in organ and tissues of occupational workers by using TLDs for the implantation of 125Ⅰ seed sources.Methods The experiments with 60Co γ-rays were carried out for the stability.A group of TLD chips was exposed to 125Ⅰ seed sources to establish standard dose curve for air kerma.During the 125Ⅰ seed implantation, the TLD chips were pasted to 13 locations like thyroid inside and outside the lead aprons worn by occupational workers to measure average absorbed dose and calculate the absorbed doses and effectives to organs and tissues.Results For 3 cases of prostate cancers with implantation of 125Ⅰ seeds, the worker's organs and tissues received the absorbed dose 0.02 -3.80 μ Gy and effective dose 0.06- 1.81 μSv outside lead aprons and the highest absorbed dose 2.35 μ Gy and effective 0.02 μSv inside lead aprons, respectively, with more than 65.9% of rays shielded.For 3 cases of brain cancers with implantation of 125Ⅰ seeds, the workers received the absorbed dose 0.23 - 11.31 μGy and effective dose 0.88 - 4.07 μSv outside lead aprons and the highest absorbed dose 2.22 μ Gy and effective dose 0.09 μSv inside lead aprons, respectively, with more than 54.5% of rays shielded.For 3 cases of lung cancers with implantation of 125Ⅰ seeds, the workers received the absorbed dose 0.03 - 14.78 μGy and effective dose 0.35 -7.59 μSv outside lead aprons and the highest absorbed dose 4.09 μGy and effective 0.22 μSv inside lead aprons, respectively, with more than 58.4% of rays shielded.For 2 cases of mediastinum cancers with implantation of 125Ⅰseeds, the workers received the absorbed dose 0.06 - 74.91 μGy and effective dose 0.83 - 17.96 μSv outside lead aprons and the highest absorbed dose 10.29 μGy and effective 0.5 μSv inside lead aprons, respectively, with more than 85% of rays shielded.For one case of ovary cancer with implantation of 125Ⅰ seeds, the worker received the absorbed dose 0.09 - 14.29 μGy and effective dose 2.40 - 4.50 μSv outside lead aprons and the highest absorbed dose 7.77 μGy and effective 0.12 μSv inside lead aprons, respectively, with more than 34% of rays shielded.For one case of eye cancer with implantation of 125Ⅰ seeds, the workers received the absorbed dose 2.2 -39.84 μGy and effective dose 4.48 - 10.06 μSv outside aprons and the highest absorbed dose 5.19 μGy and effective 0.16 μSv inside aprons, respectively, with more than 54.6 % of rays shielded.Conclusions The method of using TLDs to measure the doses to the occupational workers in the course of the implantation of 125Ⅰ seed sources is simple and easy to operate.It would be an effective approach to protecting medical workers in the case of brachytherapy.
10.Study on application of human cognition reliability model in human error in emergency response against the source blockage of high dose rate afterloading unit
Jun DENG ; Yaohua FAN ; Baorong YUE ; Kedao WEI ; Hao LI
Chinese Journal of Radiological Medicine and Protection 2011;31(4):404-407
Objective To put forward reasonable and feasible recommendations aiming at enhancing the application safety of afterloading unit, through studying the human reliability in the emergency response against the source blockage of afterloading unit.Methods Based on the human cognition reliability model, ten operation errors during the emergency response against the source blockage of afterloading unit were analyzed and permissible time widow of emergency response operation were determined.The human error probability was calculated with the execution time of emergency response operation obtained through simulation, observation and recording.Results The operation action, relevant permissible time window and execution time were obtained with the corresponding human error probabilities in the range 0.04 - 0.27.Conclusions The human error model in emergency response against the source blockage of afterloading unit based on HCRmodel is feasible, and provides important reference basis to reduce the occurrence of potential exposure and mitigate the consequence of potential exposure.