1.Long-Term Evolution of the Electrical Stimulation Levels for Cochlear Implant Patients.
Jose Luis VARGAS ; Manuel SAINZ ; Cristina ROLDAN ; Isaac ALVAREZ ; Angel DE LA TORRE
Clinical and Experimental Otorhinolaryngology 2012;5(4):194-200
OBJECTIVES: The stimulation levels programmed in cochlear implant systems are affected by an evolution since the first switch-on of the processor. This study was designed to evaluate the changes in stimulation levels over time and the relationship between post-implantation physiological changes and with the hearing experience provided by the continuous use of the cochlear implant. METHODS: Sixty-two patients, ranging in age from 4 to 68 years at the moment of implantation participated in this study. All subjects were implanted with the 12 channels COMBI 40+ cochlear implant at San Cecilio University Hospital, Granada, Spain. Hearing loss etiology and progression characteristics varied across subjects. RESULTS: The analyzed programming maps show that the stimulation levels suffer a fast evolution during the first weeks after the first switch-on of the processor. Then, the evolution becomes slower and the programming parameters tend to be stable at about 6 months after the first switch-on. The evolution of the stimulation levels implies an increment of the electrical dynamic range, which is increased from 15.4 to 20.7 dB and improves the intensity resolution. A significant increment of the sensitivity to acoustic stimuli is also observed. For some patients, we have also observed transitory changes in the electrode impedances associated to secretory otitis media, which cause important changes in the programming maps. CONCLUSION: We have studied the long-term evolution of the stimulation levels in cochlear implant patients. Our results show the importance of systematic measurements of the electrode impedances before the revision of the programming map. This report also highlights that the evolution of the programming maps is an important factor to be considered in order to determine an adequate calendar fitting of the cochlear implant processor.
Acoustics
;
Cochlear Implants
;
Electric Impedance
;
Electric Stimulation
;
Electrodes
;
Hearing
;
Hearing Loss
;
Humans
;
Otitis Media with Effusion
;
Prosthesis Fitting
;
Spain
2.White spots on the mucosal surface of the duodenum in dogs with lymphocytic plasmacytic enteritis.
Mercedes GARCIA-SANCHO ; Angel SAINZ ; Alejandra VILLAESCUSA ; Antonio RODRIGUEZ ; Fernando RODRIGUEZ-FRANCO
Journal of Veterinary Science 2011;12(2):165-169
Distended lacteals, described as expanded white villi in duodenum, are strongly indicative of primary intestinal lymphangiectasia. In the present study, we evaluated the significance of white spots present in the duodenal mucosa of dogs with lymphocytic plasmacytic enteritis (LPE). Fifty dogs with LPE were included in this study, and white spots were detected in the duodenal mucosa in 22 dogs during endoscopy. Hypoproteinemia was more frequent in dogs with white spots than in dogs without spots (p = 0.02). Serum protein and albumin concentration were significantly lower in LPE dogs with white spots (p = 0.038) compared to LPE dogs without white spots (p = 0.039). There was a significant correlation between white spots density and lymphatic dilatation histological scores (p = 0.023; rho = 0.481). These results suggest that the presence of white spots in the duodenal mucosa of dogs is not a finding exclusive for intestinal lymphangiectasia. Low serum protein and albumin concentrations together with lymphatic dilatation seem to be related to the presence of white spots in the duodenal mucosa of LPE dogs.
Animals
;
Biopsy/veterinary
;
Blood Proteins/metabolism
;
Dog Diseases/blood/*pathology
;
Dogs
;
Duodenum/*pathology
;
Endoscopy/veterinary
;
Female
;
Histocytochemistry/veterinary
;
Intestinal Mucosa/*pathology
;
Lymphangiectasis, Intestinal/blood/pathology/*veterinary
;
Male
;
Retrospective Studies
;
Statistics, Nonparametric