1.A Biomechanical Study Comparing Cerclage Wiring Performed with a Power Tool versus the Manual Method.
Kevin KOO ; Zhihong ZHOU ; Andy Ks YEW ; Joyce Sb KOH ; Tet Sen HOWE
Annals of the Academy of Medicine, Singapore 2015;44(12):554-557
INTRODUCTIONWe conducted a biomechanical study comparing cerclage wiring using a power tool with the traditional manual method.
MATERIALS AND METHODSOur study consisted of 4 experimental arms based on the method of fixation and diameter of wires. The 4 arms were: 1) power tool method using 0.8 mm cerclage wires, 2) power tool method using 1.0 mm cerclage wires, 3) conventional manual method using 0.8 mm cerclage wires, and 4) conventional manual method using 1.0 mm cerclage wires. Synthetic femur bones were employed in our study. Six specimens were prepared for each arm. Each specimen was cut lengthwise and pressure sensors were placed in between. For the power tool method, while maintaining tension, wires were coiled using the Colibri power tool until just before secondary coiling occurred. For the conventional manual method, each specimen was compressed by plier twisting for 10 rounds, while maintaining tension. Cerclaging and data recording was done thrice for each specimen, giving a total of 18 readings per arm. Peak and steady-state forces were recorded.
RESULTSThere was no significant difference between the peak forces recorded between the power drill and manual methods. The steady-state forces achieved using the power tool method were significantly higher than that achieved in the manual fixation method (0.8 mm wires: 54.89N vs 27.26N, P = 0.037; 1.0 mm wires: 71.59N vs 39.66N, P = 0.025).
CONCLUSIONThe power tool method achieved a superior steady-state force of compression across the fracture site for both 0.8 mm and 1 mm wires.
Biomechanical Phenomena ; Bone Wires ; Femur ; surgery ; Fracture Fixation, Internal ; methods ; Humans ; Models, Anatomic