1.Development and validation of a liquid chromatography-isotope dilution tandem mass spectrometry for determination of olanzapine in human plasma and its application to bioavailability study.
Mengqi ZHANG ; Jingying JIA ; Chuan LU ; Gangyi LIU ; Chengyin YU ; Yuzhou GUI ; Yun LIU ; Yanmei LIU ; Wei WANG ; Shuijun LI ; Chen YU
Acta Pharmaceutica Sinica 2010;45(6):767-71
A simple, reliable and sensitive liquid chromatography-isotope dilution mass spectrometry (LC-ID/MS) was developed and validated for quantification of olanzapine in human plasma. Plasma samples (50 microL) were extracted with tert-butyl methyl ether and isotope-labeled internal standard (olanzapine-D3) was used. The chromatographic separation was performed on XBridge Shield RP 18 (100 mm x 2.1 mm, 3.5 microm, Waters). An isocratic program was used at a flow rate of 0.4 m x min(-1) with mobile phase consisting of acetonitrile and ammonium buffer (pH 8). The protonated ions of analytes were detected in positive ionization by multiple reactions monitoring (MRM) mode. The plasma method, with a lower limit of quantification (LLOQ) of 0.1 ng x mL(-1), demonstrated good linearity over a range of 0.1 - 30 ng x mL(-1) of olanzapine. Specificity, linearity, accuracy, precision, recovery, matrix effect and stability were evaluated during method validation. The validated method was successfully applied to analyzing human plasma samples in bioavailability study.
2.Development and validation of an HPLC-MS/MS method to determine clopidogrel in human plasma.
Gangyi LIU ; Chunxia DONG ; Weiwei SHEN ; Xiaopei LU ; Mengqi ZHANG ; Yuzhou GUI ; Qinyi ZHOU ; Chen YU
Acta Pharmaceutica Sinica B 2016;6(1):55-63
A quantitative method for clopidogrel using online-SPE tandem LC-MS/MS was developed and fully validated according to the well-established FDA guidelines. The method achieves adequate sensitivity for pharmacokinetic studies, with lower limit of quantifications (LLOQs) as low as 10 pg/mL. Chromatographic separations were performed on reversed phase columns Kromasil Eternity-2.5-C18-UHPLC for both methods. Positive electrospray ionization in multiple reaction monitoring (MRM) mode was employed for signal detection and a deuterated analogue (clopidogrel-d 4) was used as internal standard (IS). Adjustments in sample preparation, including introduction of an online-SPE system proved to be the most effective method to solve the analyte back-conversion in clinical samples. Pooled clinical samples (two levels) were prepared and successfully used as real-sample quality control (QC) in the validation of back-conversion testing under different conditions. The result showed that the real samples were stable in room temperature for 24 h. Linearity, precision, extraction recovery, matrix effect on spiked QC samples and stability tests on both spiked QCs and real sample QCs stored in different conditions met the acceptance criteria. This online-SPE method was successfully applied to a bioequivalence study of 75 mg single dose clopidogrel tablets in 48 healthy male subjects.