1.Set-up uncertainties with radiation therapy for nasopharyngeal carcinoma
Cheng-Guang LIN ; Guo-Wen LI ; Lui-Wen LIN ; Wen-Jie LI ; Jun HUANG ; Jian-Xin SU ; Xiao-Wu DEN ; Nian-Ji CUI ;
Chinese Journal of Radiation Oncology 2005;0(06):-
Objective This study is to investigate the set-up accuracy of thermoplastic mask used for immobilization of nasopharyngeal carcinoma (NPC) patients being treated by Intensity Modulated Radia- tion Therapy (IMRT).Methods Nineteen patients with early stage nasopharyngeal carcinoma (T1- T2N0M0),treated by fractionated intensity-modulated radiation therapy,underwent repeated CTs during their 6-week treatment course.We evaluated their anatomic landmark coordinates in a total of 85 repeated CT data sets and respective x,y and z shifts relative to their position in the 19 treatment-planning reference CTs.Results The translation in x,y,and z-axes with their mean value derived from both positive and negative set-up errors was-0.84 mm(x-axis),+0.65 mm(y-axis) and +0.01 mm(z-axis).Mean target isocenter translation was (0.89?0.69) mm,(0.82?0.79) mm,(0.95?1.24) mm in x,y and z-direc- tions,respectively.The mean total magnitude vector and 95% CI of isocenter motion were 1.87 mm and 4.65 mm.The data measured over the 6-week fractionated course of treatment revealed a slight deterioration of isocenter accuracy.The 95% CI,considered by us to be a valuable parameter for characterizing the sys- tem,of 4.17 mm for measurement within the first 3 weeks increased to 5.12 mm in the last 3 weeks of treat- ment.Conclusions The sequential CT scanning is a pronounced valuable method of evaluating the quality of departmental specific patient positioning procedures.The thermoplastic mask is eyed as well suited tool for immobilization and repositioning of NPC patients receiving intensity-modulated radiation therapy.