1.The research on quality control parameters detection of neonatal jaundice treatment cases
Qi WANG ; Linxiang HE ; Taishi ZHENG
China Medical Equipment 2016;13(9):42-44,45
Objective: Through researching the quality control parameters detection of neonatal jaundice treatment cases, we can control the quality of treatment cases and preventive maintenance to make sure the therapy to newborn is effective and safe. Methods:Using INCU portable analyzer and BiliBlangket Light MeterII blu-ray irradiation apparatus to respectively test the temperature control parameters and the intensity of blue light phototherapy of the jaundice treatment cases, a lot of research data can be acquired. Results:On the base of researching the quality control parameters detection of neonatal jaundice treatment cases (30 sets in total), we can control the quality of treatment cases smoothly. Through these tests, we can fix the medical equipments which cannot pass the detections. Conclusion:The research on the quality control parameters detection of neonatal jaundice treatment cases makes the preventive maintenance and quality control quick and effective to ensure the treatment effect and safety for the neonatal.
2.The Resolution of Wet Wrap Bag after Disinfected in the Pulsation Sterilizer
Linxiang HE ; Qi WANG ; Taishi ZHENG
Chinese Journal of Medical Instrumentation 2016;40(6):464-465
Our hospital has instaled a couple of high temperature and high pressure pulsation sterilizers. However, after the process of disinfection and sterilization, we found that many non-woven bags in which wrapped the contaminated medical devices were wet. This abnormal phenomenon continuously has happened a couple of times. Having excessive condensed water inside the wrap bag wil severely affect the sterilization effect, so we did a scrupulous survey of the phenomenon and the structure of the machine and the steam boiler. At last, we came up with a perfect solution, that is, to make a redesign of the steam boiler and the steam conveying pipeline network.