1. Comparison of CSF analysis by manual counts and automated hematology analyzer Sysmex XN-2000
Bolor A ; Oyunkhand L ; Erdenechimeg D ; Narmandakh G ; Gerel A ; Narantuguldur D ; Oyunchimeg R ; Naran G
Health Laboratory 2016;5(1):12-15
Introduction:When a central nervous system disorder (meningitis, encephalitis, hemorrhage, leukemia infltration and other neoplasma) is present, cerebrospinal fluid (CSF) shows various changes that reflected the condition. Therefore it is essential to test CSF. Different types of CSF tests include cell count; cell differentiation; chemistry; immunology; microbiology and molecular biology. CSF cell count and cell differentiation in particular, are crucial in differentiating diagnosing various CNS disorder needing immediate care and in evaluating the treatment. The patient’s prognosis largely depends on how accurate diagnosis was done and how early treatment was provided. There for CSF test require high precision and accuracy. In Mongolia until now 2st and 3st level hospital using manual method for CSF cell count and cell differentiation test. In this test has 2 actual problems, which is depends on the analytical techniques, skills and sample stability specific problem. But in Japan in 2011 newly designed Sysmex XN Series hematology analyser with body fluid mode (CSF,pleural effusion, peritoneal and synovial fluid). On The First Central Hospital of Mongolia In 2013 frst timeinstalled Sysmex XN-2000 hematology analyser andpossible use of body fluid automatic testing methods.Materials and methods:We evaluated the basic assay performance of the body fluid mode on the automated hematology analyzer XN-2000, which is used for analysis of CSF fluid. We compared between the manual method and XN-2000 analysis for nucleated (WBC), mononuclear (MN) and polymorphonuclear (PMN) cells was also randomly studied using 10 CSF samples of inpatient section our hospital.Results:In CSF samples the coeffcient correlation(r) for WBC/µl, MN%, PMN% were respectively 0.83, 0.95 ба 0.95.Discussion:The correlation for MN%, PMN% were between automate and manual method was good, that is similar to the other researchers. Whereas the correlation for WBC/µl slightly low, this was probably correlation relatively weak or show discrepancies. In introduction inscriptive in analysis accuracy can to affect analytical techniques skills, sample stability and specifc many problems. Therefore scientifc studied and proven ability specifcity, sensitivity, reproducibility, quality, personnel low cost and spend less time, automatically Sysmex XN series hematology analyzer is desirable to domesticate an appropriate level of medical laboratories.
2.Effects of Storage Conditions on Complete Blood Cell Count Parameters
Batchimeg N ; Oyunkhand L ; Altankhuyag E ; Gantulga D ; Uranbaigali E ; Munkhtulga L
Health Laboratory 2020;11(1):18-23
Introduction:
The complete blood count (CBC) is a frequently performed laboratory test today. This study evaluated the effects of temperature and sample storage time on parameters of CBC which could produce misleading results of clinical significance.
Methods:
In a cross-sectional study, CBC was checked in 20 randomly selected out-patients and baseline measurements were analyzed using the XN-2000 (Sysmex, Kobe, Japan) fully automated hematology analyzer. CBC was done all samples of storage at room temperature. Values were checked at time intervals of 0, 6, and 24 hr.
Results:
Among CBC parameters, white blood cell, red blood cell, hemoglobin, mean cell hemoglobin (MCH), neutrophils and lymphocytes were stable at time up to 6 h. Hematocrit increased between 0 and 24 hours, averaging 41.5% and 45.2%, respectively. MCV, RDW-SD, and RDW-CV increased between 0 and 24 hours. The mean value was statistically significant. There were 85.6fL/ 93.4fL (p<0.001), 40.7fL /48.2fL (p<0.001), 13.1% and 14.2% (p<0.05), respectively.
However, the MCHC was affected by time differences. (p <0.001 at 0 and 24 hours, p <0.001 at 3 and 24 hours). Platelet PDW, MPV, and P-LCR values increased between 0 and 24 h, respectively.
Conclusion
Whole blood samples were stored at room temperature for 24 hours for CBC tests, there were statistically significant differences in the size of red blood cells and platelets.