2.Residual Limb Volume Change in the Early Post-operative Phase after Below-the-knee Amputation : A Report of Three Cases
Yohei Tanaka ; Takaaki Ueno ; Kiyokazu Tanaka ; Nobuhiko Haga
The Japanese Journal of Rehabilitation Medicine 2017;54(11):889-893
For lower limb amputees, good prosthetic fittings are important for wearing prostheses while ensuring gait stability, without skin breakdown. Poor prosthetic fittings tend to occur in the early stage after amputation because of significant changes in residual limb volume. We measured the extent of change in residual limb volume in three below-the-knee amputees by using computed tomography. The measurements were performed before and after inpatient rehabilitation for the first prosthesis. The measurement showed a remarkable change in residual limb volume in a highly active amputee without complications. In contrast, the change in residual limb volume was small in two less active below-the-knee amputees with serious complications, such as heart and renal failures. Generally, to maintain good prosthetic fittings, the first prostheses should be made during inpatient rehabilitation to facilitate maturation of the residual limbs. For the less active below-knee amputees, the prostheses could be made in the outpatient settings because the volume fluctuations of their residual limbs are small and the functional requirements for their daily living are modest.
3.Planning evaluation of stereotactic magnetic resonance–guided online adaptive radiosurgery for kidney tumors close to the organ at risk: is it valuable to wait for good timing to perform stereotactic radiosurgery?
Takaya YAMAMOTO ; Shohei TANAKA ; Noriyoshi TAKAHASHI ; Rei UMEZAWA ; Yu SUZUKI ; Keita KISHIDA ; So OMATA ; Kazuya TAKEDA ; Hinako HARADA ; Kiyokazu SATO ; Yoshiyuki KATSUTA ; Noriyuki KADOYA ; Keiichi JINGU
Radiation Oncology Journal 2025;43(1):40-48
Purpose:
This study aimed to investigate changes in target coverage using magnetic resonance–guided online adaptive radiotherapy (MRgoART) for kidney tumors and to evaluate the suitable timing of treatment.
Materials and Methods:
Among patients treated with 3-fraction MRgoART for kidney cancer, 18 tumors located within 1 cm of the gastrointestinal tract were selected. Stereotactic radiosurgery planning with a prescription dose of 26 Gy was performed using pretreatment simulation and three MRgoART timings with an adapt-to-shape method. The best MRgoART plan was defined as the plan achieving the highest percentage of planning target volume (PTV) coverage of 26 Gy. In clinical scenario simulation, MRgoART plans were evaluated in the order of actual treatment. Waiting for the next timing was done when the PTV coverage of 26 Gy did not achieve 95%–99% or did not increase by 5% or more compared to the pretreatment plan.
Results:
The median percentages of PTV receiving 26 Gy in pretreatment and the first, second, and third MRgoART were 82% (range, 19%), 63% (range, 7% to 99%), 88% (range, 31% to 99%), and 95% (range, 3% to 99%), respectively. Comparing pretreatment simulation plans with the best MRgoART plans showed a significant difference (p = 0.025). In the clinical scenario simulation, 16 of the 18 planning series, including nine plans with 95%–99% PTV coverage of 26 Gy and seven plans with increased PTV coverage by 5% or more, would be irradiated at a good timing.
Conclusion
MRgoART revealed dose coverage differences at each MRgoART timing. Waiting for optimal irradiation timing could be an option in case of suboptimal timing.
4.Planning evaluation of stereotactic magnetic resonance–guided online adaptive radiosurgery for kidney tumors close to the organ at risk: is it valuable to wait for good timing to perform stereotactic radiosurgery?
Takaya YAMAMOTO ; Shohei TANAKA ; Noriyoshi TAKAHASHI ; Rei UMEZAWA ; Yu SUZUKI ; Keita KISHIDA ; So OMATA ; Kazuya TAKEDA ; Hinako HARADA ; Kiyokazu SATO ; Yoshiyuki KATSUTA ; Noriyuki KADOYA ; Keiichi JINGU
Radiation Oncology Journal 2025;43(1):40-48
Purpose:
This study aimed to investigate changes in target coverage using magnetic resonance–guided online adaptive radiotherapy (MRgoART) for kidney tumors and to evaluate the suitable timing of treatment.
Materials and Methods:
Among patients treated with 3-fraction MRgoART for kidney cancer, 18 tumors located within 1 cm of the gastrointestinal tract were selected. Stereotactic radiosurgery planning with a prescription dose of 26 Gy was performed using pretreatment simulation and three MRgoART timings with an adapt-to-shape method. The best MRgoART plan was defined as the plan achieving the highest percentage of planning target volume (PTV) coverage of 26 Gy. In clinical scenario simulation, MRgoART plans were evaluated in the order of actual treatment. Waiting for the next timing was done when the PTV coverage of 26 Gy did not achieve 95%–99% or did not increase by 5% or more compared to the pretreatment plan.
Results:
The median percentages of PTV receiving 26 Gy in pretreatment and the first, second, and third MRgoART were 82% (range, 19%), 63% (range, 7% to 99%), 88% (range, 31% to 99%), and 95% (range, 3% to 99%), respectively. Comparing pretreatment simulation plans with the best MRgoART plans showed a significant difference (p = 0.025). In the clinical scenario simulation, 16 of the 18 planning series, including nine plans with 95%–99% PTV coverage of 26 Gy and seven plans with increased PTV coverage by 5% or more, would be irradiated at a good timing.
Conclusion
MRgoART revealed dose coverage differences at each MRgoART timing. Waiting for optimal irradiation timing could be an option in case of suboptimal timing.
5.Planning evaluation of stereotactic magnetic resonance–guided online adaptive radiosurgery for kidney tumors close to the organ at risk: is it valuable to wait for good timing to perform stereotactic radiosurgery?
Takaya YAMAMOTO ; Shohei TANAKA ; Noriyoshi TAKAHASHI ; Rei UMEZAWA ; Yu SUZUKI ; Keita KISHIDA ; So OMATA ; Kazuya TAKEDA ; Hinako HARADA ; Kiyokazu SATO ; Yoshiyuki KATSUTA ; Noriyuki KADOYA ; Keiichi JINGU
Radiation Oncology Journal 2025;43(1):40-48
Purpose:
This study aimed to investigate changes in target coverage using magnetic resonance–guided online adaptive radiotherapy (MRgoART) for kidney tumors and to evaluate the suitable timing of treatment.
Materials and Methods:
Among patients treated with 3-fraction MRgoART for kidney cancer, 18 tumors located within 1 cm of the gastrointestinal tract were selected. Stereotactic radiosurgery planning with a prescription dose of 26 Gy was performed using pretreatment simulation and three MRgoART timings with an adapt-to-shape method. The best MRgoART plan was defined as the plan achieving the highest percentage of planning target volume (PTV) coverage of 26 Gy. In clinical scenario simulation, MRgoART plans were evaluated in the order of actual treatment. Waiting for the next timing was done when the PTV coverage of 26 Gy did not achieve 95%–99% or did not increase by 5% or more compared to the pretreatment plan.
Results:
The median percentages of PTV receiving 26 Gy in pretreatment and the first, second, and third MRgoART were 82% (range, 19%), 63% (range, 7% to 99%), 88% (range, 31% to 99%), and 95% (range, 3% to 99%), respectively. Comparing pretreatment simulation plans with the best MRgoART plans showed a significant difference (p = 0.025). In the clinical scenario simulation, 16 of the 18 planning series, including nine plans with 95%–99% PTV coverage of 26 Gy and seven plans with increased PTV coverage by 5% or more, would be irradiated at a good timing.
Conclusion
MRgoART revealed dose coverage differences at each MRgoART timing. Waiting for optimal irradiation timing could be an option in case of suboptimal timing.
6.Planning evaluation of stereotactic magnetic resonance–guided online adaptive radiosurgery for kidney tumors close to the organ at risk: is it valuable to wait for good timing to perform stereotactic radiosurgery?
Takaya YAMAMOTO ; Shohei TANAKA ; Noriyoshi TAKAHASHI ; Rei UMEZAWA ; Yu SUZUKI ; Keita KISHIDA ; So OMATA ; Kazuya TAKEDA ; Hinako HARADA ; Kiyokazu SATO ; Yoshiyuki KATSUTA ; Noriyuki KADOYA ; Keiichi JINGU
Radiation Oncology Journal 2025;43(1):40-48
Purpose:
This study aimed to investigate changes in target coverage using magnetic resonance–guided online adaptive radiotherapy (MRgoART) for kidney tumors and to evaluate the suitable timing of treatment.
Materials and Methods:
Among patients treated with 3-fraction MRgoART for kidney cancer, 18 tumors located within 1 cm of the gastrointestinal tract were selected. Stereotactic radiosurgery planning with a prescription dose of 26 Gy was performed using pretreatment simulation and three MRgoART timings with an adapt-to-shape method. The best MRgoART plan was defined as the plan achieving the highest percentage of planning target volume (PTV) coverage of 26 Gy. In clinical scenario simulation, MRgoART plans were evaluated in the order of actual treatment. Waiting for the next timing was done when the PTV coverage of 26 Gy did not achieve 95%–99% or did not increase by 5% or more compared to the pretreatment plan.
Results:
The median percentages of PTV receiving 26 Gy in pretreatment and the first, second, and third MRgoART were 82% (range, 19%), 63% (range, 7% to 99%), 88% (range, 31% to 99%), and 95% (range, 3% to 99%), respectively. Comparing pretreatment simulation plans with the best MRgoART plans showed a significant difference (p = 0.025). In the clinical scenario simulation, 16 of the 18 planning series, including nine plans with 95%–99% PTV coverage of 26 Gy and seven plans with increased PTV coverage by 5% or more, would be irradiated at a good timing.
Conclusion
MRgoART revealed dose coverage differences at each MRgoART timing. Waiting for optimal irradiation timing could be an option in case of suboptimal timing.
7.Planning evaluation of stereotactic magnetic resonance–guided online adaptive radiosurgery for kidney tumors close to the organ at risk: is it valuable to wait for good timing to perform stereotactic radiosurgery?
Takaya YAMAMOTO ; Shohei TANAKA ; Noriyoshi TAKAHASHI ; Rei UMEZAWA ; Yu SUZUKI ; Keita KISHIDA ; So OMATA ; Kazuya TAKEDA ; Hinako HARADA ; Kiyokazu SATO ; Yoshiyuki KATSUTA ; Noriyuki KADOYA ; Keiichi JINGU
Radiation Oncology Journal 2025;43(1):40-48
Purpose:
This study aimed to investigate changes in target coverage using magnetic resonance–guided online adaptive radiotherapy (MRgoART) for kidney tumors and to evaluate the suitable timing of treatment.
Materials and Methods:
Among patients treated with 3-fraction MRgoART for kidney cancer, 18 tumors located within 1 cm of the gastrointestinal tract were selected. Stereotactic radiosurgery planning with a prescription dose of 26 Gy was performed using pretreatment simulation and three MRgoART timings with an adapt-to-shape method. The best MRgoART plan was defined as the plan achieving the highest percentage of planning target volume (PTV) coverage of 26 Gy. In clinical scenario simulation, MRgoART plans were evaluated in the order of actual treatment. Waiting for the next timing was done when the PTV coverage of 26 Gy did not achieve 95%–99% or did not increase by 5% or more compared to the pretreatment plan.
Results:
The median percentages of PTV receiving 26 Gy in pretreatment and the first, second, and third MRgoART were 82% (range, 19%), 63% (range, 7% to 99%), 88% (range, 31% to 99%), and 95% (range, 3% to 99%), respectively. Comparing pretreatment simulation plans with the best MRgoART plans showed a significant difference (p = 0.025). In the clinical scenario simulation, 16 of the 18 planning series, including nine plans with 95%–99% PTV coverage of 26 Gy and seven plans with increased PTV coverage by 5% or more, would be irradiated at a good timing.
Conclusion
MRgoART revealed dose coverage differences at each MRgoART timing. Waiting for optimal irradiation timing could be an option in case of suboptimal timing.