Objective To study the calculational method for the radiotherapy facilities of the medical linear accelerator' s useful beam towars its maze.Methods The shielding calculation was made under the relevant national standards for a radiotherapy treatment room and compared with the test results.Results The dose rates at the maze entrance as calculated and measured were 89 and 86 μSv/h inside the maze door,as well as 5.7 and 6.2 μSv/h outside the maze door,respectively.The calculated results were consistent with measured results.Conclusions By comparison of calculated results with measured results,the accuracy of the theoretical calculation method could be verified.