Development and verification of pre-column derivatization reverse phase-high performance liquid chromatography for determination of Glycine and Histidine content in recombinant proteins
10.13200/j.cnki.cjb.003947
- VernacularTitle:重组蛋白中甘氨酸、组氨酸含量柱前衍生反相高效液相色谱检测方法的建立及验证
- Author:
ZHANG Xiaoxue
- Publication Type:Journal Article
- Keywords:
Pre-column derivatization reverse phase-high performance liquid chromatography(RP-HPLC);
Glycine;
Histidine;
Content
- From:
Chinese Journal of Biologicals
2023;36(7):844-849
- CountryChina
- Language:Chinese
-
Abstract:
Objective To develop and verify a pre-column derivatization reverse phase-high performance liquid chromatography(RP-HPLC) method for determination of Glycine and Histidine content in recombinant proteins.Methods AccQ Tag-C 18(3.9 mm × 150 mm,4 μm) column was used as chromatographic column,6-aminoquinolyl-N-hydroxysccinimidyl carbamate(AQC) was used as pre-column derivatization reagent,while α-aminobutyric acid as internal standard.AccQTag Eluent A solution,acetonitrile solution and high-purity water were used as mobile phases.The UV detection wavelength was 248 nm,injection volume was 10 μL,flow rate was 1.0 mL/min,and column temperature was 37 ℃.The contents of Glycine and Histidine in samples were determined by the internal standard method,and the specificity,linearity,detection limit,quantitative limit,precision,accuracy and stability of the method were verified.Results The developed method effectively separated Glycine,Histidine and internal standard α-aminobutyric acid with high specificity.The standard curves of Glycine in the range of 2.25~11.25 μg/mL and Histidine in the range of 72.85~364.24 μg/mL showed good linearity,each correlation coefficient(R~2) 0.99.The detection limits were 2.25 μg/mL for Glycine and 18.21 μg/mL for Histidine.The quantitative limits were 4.69 μg/mL for Glycine and 32.86 μg/mL for Histidine.The relative standard deviation(RSD) of 6 replicates with the same concentration of Glycine and Histidine were 4.6% and 5.0%,and the RSD of recovery rate in intermediate precision test was 6.9% and 2.0%,respectively.The content of Glycine was close to the quantitative limit,and the average recoveries of high,medium and low concentrations of samples were within 75.9%~111.7%;The recoveries of Histidine ranged from 88.9% to 97.3%.The RSD of Glycine content and Histidine content was 7.7% and 3.3% respectively at 0,12,18,24,30 and 48 h in the same sample.Conclusion The pre-column derivatization RP-HPLC method has accurate and reliable results with high precision,which might be used for quality control of Glycine and Histidine content in recombinant proteins.