Effectiveness of S-curve leakage testing method on flexible endoscope
10.3760/cma.j.cn321463-20220922-00276
- VernacularTitle:软式内镜使用S弯精细测漏法的效果分析
- Author:
Jing LI
1
;
Na HUI
;
Hui LUO
;
Xuegang GUO
Author Information
1. 空军军医大学第一附属医院消化内科,西安 710032
- Keywords:
Endoscopes;
Maintenance;
Cost-benefit analysis
- From:
Chinese Journal of Digestive Endoscopy
2023;40(4):313-315
- CountryChina
- Language:Chinese
-
Abstract:
To investigate the effects of the S-curve leakage testing method on the detection of flexible endoscope and occurrence of medium to high-level faults of flexible endoscope. A convenience sampling method was used to study the information of endoscopic leak detection at the digestive endoscopic centre of the First Affiliated Hospital of Air Force Medical University. From July 2016 to December 2017, 58 endoscopes which received conventional leakage testing were set as the control group (29 116 tests). From January 2018 to June 2019, 56 endoscopes which received S-curve leakage testing were set as the observation group (28 112 tests). The results showed that the detection proportion of angular abnormalities in the observation group was higher than that in the control group [65.6% (59/90) VS 40.4% (36/89), χ2=11.330, P<0.001]. However, the detection proportion of medium to high-grade faults such as damaged charge coupled device (CCD) components and broken insertion section in the observation group was lower than that in the control group [0.04% (10/28 112) VS 0.08% (23/29 116), χ2=4.680, P=0.030]. Moreover, the overall maintenance cost was lower in the observation group than that in the control group (313.7 thousand yuan VS 942.6 thousand yuan). It is indicated that the S-curve leakage testing method can detect low-level endoscopic faults early such as angular abnormalities, and reduce the occurrence of medium to high-level endoscopic faults, which contributes to cost reduction and efficiency increase.