Value of short-time transcutaneous carbon dioxide and oxygen pressure monitoring in critically ill preterm infants
10.3760/cma.j.cn113903-20230216-00070
- VernacularTitle:经皮二氧化碳分压和经皮氧分压短时监测在重症早产儿中的应用价值
- Author:
Yuejyu CAI
1
;
Xiaolan LI
;
Xiao RONG
;
Weineng LU
;
Huayan ZHANG
;
Xiaopeng ZHAO
Author Information
1. 广州市妇女儿童医疗中心新生儿科,广州 510623
- Keywords:
Blood gas analysis;
Infant, premature;
Blood gas monitoring, transcutaneous;
Intensive care, neonatal
- From:
Chinese Journal of Perinatal Medicine
2023;26(8):658-663
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To investigate the value of short-time transcutaneous carbon dioxide pressure (TcPCO 2) and transcutaneous oxygen pressure (TcPO 2) monitoring in critically ill preterm infants. Methods:From January to December 2018, 62 critically ill neonates receiving respiratory support at Guangzhou Women and Children's Medical Center were retrospectively enrolled. A total of 348 sets of paired data including TcPCO 2/TcPO 2 and arterial carbon dioxide pressure (PaCO 2)/arterial oxygen partial pressure (PaO 2) were analyzed. The patients were divided into different groups based upon birth weight (23 cases>1 000 g-≤1 500 g, 129 sets of paired data; 18 cases≤1 000 g, 130 sets of paired data) and gestational age (16 cases born at ≤28 gestational weeks, 127 sets of paired data; 29 cases born at 28-34 gestational weeks, 159 sets of paired data) and the differences between groups were compared. The correlation and consistency of TcPCO 2/TcPO 2 and PaCO 2/PaO 2 were evaluated using Pearson's correlation and Blan-Altman scatter plots. Receiver operating characteristic (ROC) curve was drawn to analyze the diagnostic efficacy of TcPCO 2 in neonates with hypercapnia. Results:There was a positive correlation between TcPCO 2 and PaCO 2 in all patients ( r=0.913, 95% CI:0.894-0.929, P<0.05). In patients whose birth weight was>1 000 g-≤1 500 g or≤1 000 g, TcPCO 2 and PaCO 2 were positively correlated and the consistency were good ( r=0.909, 95% CI:0.874-0.935; r=0.934, 95% CI:0.908-0.953; both P<0.05), and the same finding was also observed in patients born at≤28 gestational weeks or 28-34 weeks of gestation ( r=0.938, 95% CI:0.913-0.956; r=0.871, 95% CI: 0.827-0.904; both P<0.05). The sensitivity, specificity and area under curve of TcPCO 2 in the diagnosis of hypercapnia were 90.91%, 85.85%, and 0.942, respectively. There was a poor correlation between TcPO 2 and PaO 2 in all patients and those with birth weight >1 000 g-≤1 500 g or gestational age 28-34 weeks (all r<0.75, all P<0.05). There was no correlation between TcPO 2 and PaO 2 in the birth weight ≤1 000 g and gestational age ≤28 weeks groups (both P>0.05). Conclusions:Short-time TcPCO 2 monitoring can accurately assess PaCO 2 in critically ill neonates requiring respiratory support and is of high diagnostic value for hypercapnia. However, TcPO 2 has limitation in evaluating PaO 2 and other indicators may need to be involved.