Safety of leadless pacemaker implantation in super-elderly patients
10.3760/cma.j.cn114798-20221229-01194
- VernacularTitle:超高龄患者植入无导线起搏器的安全性初探
- Author:
Jinshan HE
1
;
Jiangbo DUAN
;
Sicong LI
;
Long WANG
;
Ding LI
;
Feng ZE
;
Cuncao WU
;
Xu ZHOU
;
Cuizhen YUAN
;
Xuebin LI
Author Information
1. 北京大学人民医院心内科,北京 100044
- Keywords:
Pacemaker, artificial;
Oldest old;
Safety;
Leadless pacemaker
- From:
Chinese Journal of General Practitioners
2023;22(4):399-403
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To evaluate the safety of leadless pacemaker implantation in super-elderly patients.Methods:Eleven patients with average age of 90 (86, 92) years who underwent leadless pacemaker implantation in the Department of Cardiology, Peking University People′s Hospital from March 2021 to May 2022 were included. The clinical data and implantation information were collected. The complications (cardiac tamponade, myocardial infarction, cerebral infarction, pulmonary embolism, pacemaker reinfection, femoral vein hematoma) and death of patients were documented at 24 h, 3 d, and 1, 3, 6 months after pacemaker implantation.Results:There were 9 males and 2 females with the body mass index of 21(19, 23)kg/m 2. The underlying diseases were hypertension, diabetes, coronary heart disease, chronic kidney disease, chronic obstructive pulmonary disease, previous cerebral infarction and moderate to severe tricuspid regurgitation in 9, 9, 9, 6, 4, 4, 4 patients, respectively. The left ventricular ejection fraction was 49% (45%, 52%), the hemoglobin concentration was 118 (114, 122)g/L, 4 patients were taking anticoagulant drugs, and 6 patients were taking antiplatelet drugs. Eight patients were newly implanted with a leadless pacemaker, 2 patients were implanted after removal the old ones, and 1 case was implanted at the same time as removal. The implantation time was 45(40, 47) minutes, the X-ray exposure time was 14 (13, 15) minutes, the release time was 1 (1, 2), the threshold value was 0.50(0.38, 0.75)V/0.24 ms, the impedance was 730 (700, 770) Ω, and the perceived R-wave amplitude 8.2(6.7, 12.8) mV. During the follow-up period of 8 (6, 10) months, no patient had pacemaker dysfunction; and the threshold, R wave sensing, and impedance were stable and maintained within the normal range. No cardiac tamponade, myocardial infarction, cerebral infarction, pulmonary embolism, pacemaker reinfection or death occurred perioperatively and during the follow-up period; 1 patient had hematoma after femoral vein puncture, which improved after compression treatment. Conclusion:This single-center and small-sample study shows that leadless pacemaker implantation is safe for super-elderly patients.