Added value of T 1-weighted StarVIBE sequence for PET/MR image quality
10.3760/cma.j.cn321828-20211014-00353
- VernacularTitle:T 1加权StarVIBE序列对PET/MR图像质量的提升价值
- Author:
Hongping MENG
1
;
Xinyun HUANG
;
Xiaoyue CHEN
;
Rui GUO
;
Xiaozhu LIN
;
Jin WANG
;
Biao LI
;
Miao ZHANG
Author Information
1. 上海交通大学医学院附属瑞金医院核医学科,上海 200025
- Keywords:
Image processing, computer-assisted;
Positron-emission tomography;
Tomography, X-ray computed;
Fluorodeoxyglucose F18;
Magnetic resonance imaging
- From:
Chinese Journal of Nuclear Medicine and Molecular Imaging
2023;43(3):156-160
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To explore the added value of T 1-weighted stack-of-stars volumetric interpolated body examination (StarVIBE) sequence on PET/MR image quality. Methods:A retrospective analysis was performed on 60 patients (42 males, 18 females; age 11-86 (58±12) years) who underwent 18F-FDG PET/MR examination and with positive PET results in Ruijin Hospital, Shanghai Jiao Tong University School of Medicine from April 2020 to April 2021. All patients completed StarVIBE sequence collection, and volumetric interpolated body examination (VIBE) sequence was used as control. StarVIBE and VIBE sequence images were evaluated independently using five-point method by two physicians. The evaluation was carried out from six aspects: lesion display, lesion boundary display, vascular around lesions display, fusion level with PET image, image artifact and overall image quality. Wilcoxon signed rank test was used to compare the image quality of the two sequences, and Kappa test was performed to assess the consistency of the image quality scores between the two physicians. Results:There were 26 cases with cervical lesions, 14 cases with chest lesions, 7 cases with abdomen lesions and 13 cases with pelvic lesions. The scores of lesion display (4.0(3.8, 4.5) vs 3.5(3.0, 4.0)), lesion boundary display (4.0(4.0, 4.0) vs 3.0(3.0, 3.5)), vascular around lesions display (5.0(4.0, 5.0) vs 4.0(3.5, 4.5)), fusion level with PET image (5.0(5.0, 5.0) vs 4.5(4.0, 5.0)), image artifact (4.5(4.0, 5.0) vs 4.5(4.0, 5.0)) and overall image quality (5.0(4.0, 5.0) vs 4.0(4.0, 4.0)) of StarVIBE sequences were better than those of VIBE sequences ( z values: 3.77-6.54, all P<0.001). On the vascular around the lesions display, the scores of StarVIBE were significantly better than those of VIBE sequence in the neck (5.0(4.5, 5.0) vs 3.0(2.7, 3.5); z=4.49, P<0.001) and chest (4.5(4.3, 4.7) vs 4.0(3.6, 4.3); z=3.10, P=0.002). As for image quality, the scores of StarVIBE were also significantly better than those of VIBE in neck (5.0(4.5, 5.0) vs 4.0(3.7, 4.5); z=4.36, P<0.001) and chest (5.0(5.0, 5.0) vs 4.0(4.0, 4.5); z=3.02, P=0.003). In abdominal lesions, the score of StarVIBE was higher than that of VIBE in blood vessels (4.5(3.5, 5.0) vs 4.0(3.5, 4.5); z=2.07, P=0.038), and there was no difference between score of overall image quality (4.0(3.7, 4.5) vs 4.0(3.5, 4.5); z=0.27, P=0.785). The score of overall image quality of pelvic StarVIBE sequence was better than that of VIBE sequence (5.0(4.5, 5.0) vs 4.0(4.0, 4.5); z=2.12, P=0.034). Kappa value of image quality score between two physicians was 0.554, indicating moderate consistency. Conclusion:In whole-body PET/MR imaging, StarVIBE sequence can significantly improve the image quality of cervical, thoracic and pelvic lesions when comparing with VIBE sequence.