Application of ANSYS Fatigue Analysis in Type Selection and Evaluation of Bone Plate
10.16156/j.1004-7220.2023.03.19
- VernacularTitle:ANSYS 疲劳分析在接骨板选型评估中的应用
- Author:
Shurong LI
1
;
Shiyun LI
1
;
Ying XIONG
2
;
Ziyi HUANG
2
;
Xiaoping YANG
3
Author Information
1. Faculty of Mechanical and Electrical Engineering, Kunming University of Science and Technology
2. Department of Orthopedics, the Affiliated Yan’ an Hospital of Kunming Medical University
3. Yunnan Opax Medical Technology Co. , Lt.
- Publication Type:Journal Article
- Keywords:
bone plate;
fatigue service life;
bending strength;
fatigue analysis
- From:
Journal of Medical Biomechanics
2023;38(3):E556-E560
- CountryChina
- Language:Chinese
-
Abstract:
Objective By comparing the fatigue strength of type A and type B locking compression plates (LCP) in distal femoral plate, a theoretical evaluation method was provided for type selection of bone plate when testing its bending strength and fatigue performance. Methods Through bending strength performance test and fatigue performance test on bone plates with different types, combined with ANSYS Workbench, the finite element analysis on total deformation, von Mises stress and fatigue service life of bone plates were conducted. Results The fatigue strength of type A plate was 30.7% higher than that of type B plate, the stress of type A plate was lower than that of type B plate, and the minimum fatigue service life of type A plate was 17% higher than that of type B plate. Conclusions The fatigue performance of type A plate is better than that of type B plate, so the failure possibility of type A plate was lower than that of type B plate.The results provide references for assisting selection of different bone plates when testing the performance of two newly developed bone plates.