Agreement of corneal high-order aberrations measured by three wavefront aberrometers in myopic adults
10.3980/j.issn.1672-5123.2023.10.23
- VernacularTitle:三种波前像差仪测量成年近视患者角膜高阶像差的一致性
- Author:
Hua-Xian ZOU
1
;
Qi CHEN
1
;
Li-Li LI
1
;
Die-Feng WEI
1
;
He-Juan MAO
1
;
Yan-Yan HUANG
1
;
Peng-Fei LU
1
;
Hai-Bin ZHONG
1
Author Information
1. Guilin Medical University, Guilin 541199, Guangxi Zhuang Autonomous Region, China
- Publication Type:Journal Article
- Keywords:
Topcon KR-1W;
i.Profiler;
OPD-Scan Ⅲ;
corneal aberration;
agreement
- From:
International Eye Science
2023;23(10):1723-1731
- CountryChina
- Language:Chinese
-
Abstract:
AIM:To evaluate the agreement of corneal high-order aberrations from Topcon KR-1W, i.Profiler and OPD-Scan Ⅲ wavefront aberrometers in myopic adults.METHODS:A prospective clinical study. A total of 92 adult patients(92 eyes)with myopia in the department of optometry, the People's Hospital of Guangxi Zhuang Autonomous Region from June to August 2022 were enrolled. The third-order and fourth-order corneal aberrations at the pupil diameter of 4 and 6mm were measured by Topcon KR-1W, i.Profiler, and OPD-Scan Ⅲ, respectively. The difference and agreement of the three aberrometers were evaluated.RESULTS: The measurements at 6mm pupil diameter were all greater than those at 4mm pupil diameter. Although there were no statistical differences in the measurements of Z-44、Z-24 by the three aberrometers at 4 pupil diameter(P>0.05), there were statistical differences in other measurements(P<0.05). The aberration results measured by the three aberrometers were statistically different at the 6mm pupil diameter(P<0.05). The 95% limit of agreement(95%LoA)of the measurements of higher-order aberration, including the third-order aberrations at 4mm pupil diameter and the third-order and fourth-order aberrations at 6mm pupil diameter(except for the Z-24)were greater than 0.1μm. The concordance correlation coefficient(Pc)was lower than 0.90, indicating a poor consistency. The correlation coefficients of corneal higher-order aberrations were significantly different among the three aberrometers at 4 and 6mm pupil diameter(r4mm=0.215~0.805, P4mm<0.05; r6mm=0.561~0.916, P6mm<0.001).CONCLUSION:There were significant differences in the measurements of the third- and fourth-order corneal aberrations at 4 and 6mm pupil diameter among Topcon KR-1W, i.Profiler, and OPD-Scan Ⅲ, and the agreements were poor, so they are not interchangeably in clinical applications.