Comparison of thresholds acquired with SVR and PTA in normal hearing subjects.
- Author:
Hui LIU
1
;
Guang-you ZHU
;
Xiao-ping YANG
;
Xiao-rong ZHOU
;
Lei WAN
;
Li-hua FAN
Author Information
1. Department of Forensic Medicine, Medical College of Soochow University, Suzhou 215123, China. hhcomeon@126.com
- Publication Type:Journal Article
- MeSH:
Acoustic Stimulation;
Adult;
Analysis of Variance;
Audiometry, Evoked Response/methods*;
Audiometry, Pure-Tone;
Auditory Cortex/physiology*;
Auditory Threshold/physiology*;
Evoked Potentials, Auditory;
Female;
Hearing Disorders/diagnosis*;
Humans;
Male;
Middle Aged;
Reaction Time;
Young Adult
- From:
Journal of Forensic Medicine
2010;26(1):18-21
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVE:To investigate the value of slow vertex response (SVR) for forensic appraisement of hearing dysfunction by comparing the thresholds acquired with SVR and pure tone audiometry (PTA).
METHODS:Forty-six subjects with normal hearing were tested with PTA (0.125-8 kHz) and SVR (0.5-4 kHz). Paired t-test analysis was performed on the thresholds of SVR and PTA at each frequency (0.5, 1, 2 and 4 kHz), and analysis of variance was performed on the difference between the pure tone audiometry threshold and the evoked potential threshold among different frequency.
RESULTS:At each frequency(0.5, 1, 2 kHz and 4 kHz), difference between the thresholds acquired with SVR and PTA was statistically significant(P < 0.05). The mean difference of the two thresholds were (5.98 +/- 6.72) dB HL at 0.5 kHz, (3.86 +/- 6.12) dB HL at 1 kHz, (7.12 +/- 6.56) dB HL at 2 kHz, (8.53 +/- 7.90) dB HL at 4kHz, and the mean difference at 1 kHz was the smallest. This variation between frequencies was also statistically significant.
CONCLUSION:The thresholds acquired with SVR were higher than those acquired with PTA. The adjusted thresholds should be established at each frequency, when the SVR thresholds are used to evaluate the behavioural hearing thresholds.