Simulator of Heart Rate and Respiratory Rate for Monitoring Accuracy of Intelligent Sleep Monitoring Devices.
10.3969/j.issn.1671-7104.2023.04.018
- Author:
Li FU
1
;
Shuo JIANG
1
;
Jia TIAN
1
;
Feifei TAN
1
;
Tiantian CHEN
1
;
Zaiai ZHANG
1
;
Xiyan DING
1
;
Ke ZHANG
1
Author Information
1. Shandong Institute of Medical Device and Pharmaceutical Packaging Inspection, Jinan, 250101.
- Publication Type:Journal Article
- Keywords:
accuracy;
heart rate and respiratory rate simulator;
pressure sensors;
test;
the intelligent sleep monitoring devices
- From:
Chinese Journal of Medical Instrumentation
2023;47(4):445-448
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVE:Invent a simulator which provides a simulation of heart rate and respiratory rate to the intelligent sleep monitoring devices based on precision pressure sensors.
METHODS:The simulator was composed of control part and simulated silicone doll. The simulated silicone doll contains heartbeat simulator and breathing simulation airbag. Heartbeat and breathing combination pressure signal can be produced according to frequency set values. Frequencies of pressure signal of the simulator were compared with the monitoring results of intelligent sleep monitoring devices with known accuracy to verify the frequency accuracy of pressure signal of the simulator. Verified the repeatability and stability of the simulator with a stopwatch.
RESULTS:The heart rate of the simulator was with in ±2 beats per minute of the monitoring results of intelligent sleep monitoring devices and the respiratory rate of the simulator was with in ±2 times per minute of the monitoring results. The repeatability and stability of the simulator was better than ±5% according to results with a stopwatch.
CONCLUSIONS:It's practicable to use the simulator which provides a simulation of heart rate and respiratory rate to the accuracy test of the intelligent sleep monitoring devices based on precision pressure sensor.