Application and evaluation of droplet digital polymerase chain reaction in detection of severe acute respiratory syndrome corona virus-2
- Author:
CHEN Shuang
;
WANG Ming-yue
;
ZU Zhen
;
TANG Yun
;
YE Sheng
;
LING Hua
;
TAN Zhang-ping
- Publication Type:Journal Article
- Keywords:
Severe acute respiratory syndrome corona virus-2;
droplet digital polymerase chain reaction;
sensitivity;
specificity;
repeatability
- From:
China Tropical Medicine
2023;23(5):495-
- CountryChina
- Language:Chinese
-
Abstract:
Abstract: Objective To analyze the nucleic acid detection results of severe acute respiratory syndrome corona virus-2 (SARS-CoV-2) by droplet digital PCR (ddPCR) and compare with the detection results of real-time fluorescence quantitative RT-PCR (qRT-PCR), so as to evaluate the advantages and disadvantages of detection, and to provide data support for optimizing the nucleic acid detection scheme of SARS-CoV-2. Methods According to the SARS-CoV-2 specific primer probe published by the China Center for Disease Control and Prevention, a ddPCR detection method for SARS-CoV-2 was designed. One sample was selected for sensitivity test after gradient dilution; six respiratory virus nucleic acid positive samples including seasonal H3N2 influenza virus and SARS-CoV-2 positive samples were selected for specificity test; five SARS-CoV-2 positive samples were selected for repeatability test; in addition, 30 positive and 20 negative SARS-CoV-2 samples were selected for multiple clinical samples testing, and the results were analyzed and compared with those of qRT-PCR. Results The ddPCR method can specifically detect SARS-CoV-2, and directly obtain the original copy number of the sample target gene to achieve accurate quantification; the sensitivity test of gradient dilution positive samples showed that qRT-PCR detected target genes in part of the 10-5 dilution of samples, and no target genes were detected in 10-6 dilution, while ddPCR detected all target genes in both 10-5 and 10-6 dilution of samples. The detection limit of ddPCR was two orders of magnitude higher than that of qRT-PCR, and the sensitivity was higher than that of qRT-PCR; in the comparison of the repeatability test results of the two methods, the coefficient of variation of ddPCR was 1.266%-11.814%, lower than 1.729%-26.174% of qRT PCR, and the repeatability was higher than qRT-PCR; among 50 clinical samples, 30 positive samples of confirmed cases of Coronavirus Disease 2019 (COVID-19) were detected by both methods, SARS-CoV-2 was successfully detected by both methods, and 20 negative samples of COVID-19 were detected by both methods, and the results were negative, with a coincidence rate of 100.00% (50/50). Conclusion The ddPCR method can accurately quantify SARS-CoV-2 with strong specificity, and its sensitivity and repeatability are higher than those of qRT-PCR, but it also has certain detection limitations and is more suitable for the detection of low load samples. In the actual detection, the two methods can be reasonably combined to improve the detection accuracy.
- Full text:10.Application and evaluation of droplet digital polymerase chain reaction in detection of severe acute respiratory syndrome corona virus-2.pdf