Application of 70 kVp tube voltage combined with the FLASH in low-dose CT of pediatric paranasal sinus
10.13491/j.issn.1004-714X.2021.04.022
- VernacularTitle:超低管电压(70 kVp)联合FLASH技术在儿童鼻旁窦CT中的应用
- Author:
Jing CHI
1
;
Dongfeng XU
1
;
Shengnan YIN
1
;
Ning DING
1
;
Xuewen ZHAO
1
;
Yiding JI
1
Author Information
1. Department of Radiology, Suzhou Ninth People's Hospital, Suzhou 215200 China.
- Publication Type:SupervisionandManagement/OriginalArticles
- Keywords:
Dual-source CT;
Image Quality;
Radiation Dose;
High-pitch Protocol;
Sinusitis
- From:
Chinese Journal of Radiological Health
2021;30(4):501-505
- CountryChina
- Language:Chinese
-
Abstract:
Objective To evaluate the feasibility of dual-source low-dose computed tomography (CT) of the paranasal sinus in children, with acquisition at an ultra-low tube voltage (70 kVp) combined with the Flash scan. Methods CT scans of the paranasal sinus were performed on 80 pediatric patients who were divided into two groups according to different protocols (70 kVp protocol with Flash scan mode and the iterative reconstruction, pitch 3, the experimental group (group A), n = 40; 80 kVp protocol with conventional spiral mode, pitch 1.5, control group (group B), n = 40). For each examination, the CT dose index (CTDIvol), dose-length product (DLP) and effective dose (ED) were estimated. The image noise, signal-to-noise ratio (SNR) and overall subjective diagnostic image quality were also evaluated. Results The images of these two groups were all satisfied for the clinical diagnosis. For radiation dose, the CTDIvol (mGy), DLP (mGycm) and ED (mSv) values of the 70 kVp protocol were significantly lower than those of the 80 kVp protocol [CTDIvol: 0.39 ± 0.004 vs1.57 ± 0.009 mGy, P < 0.001; DLP: 6.31 ± 0.52 vs 19.88 ± 2.01 mGycm, P < 0.001; ED: 0.024 ± 0.005 vs 0.079 ± 0.016 mSv, P < 0.001. Compared with those of the 80-kVp protocol, the image noise and the SNRbone increased, the SNRsoft-tissue decreased. There was no statistical difference in the subjective scores of the two groups of image quality by the two physicians (P > 0.05). Conclusion When diagnosing the paranasal sinus in children, an ultralow tube voltage (70 kVp) combined with the Flash scan technique can reduce the radiation dose significantly while maintaining diagnostic image quality with clinically acceptable image noise.