Difference and consistency of corneal refractive power and astigmatism in cataract patients measured by CASIA2 and IOL Master 700
10.3980/j.issn.1672-5123.2023.6.27
- VernacularTitle:CASIA2与IOL Master 700测量白内障患者角膜屈光力及散光的差异性和一致性
- Author:
Lu-Sha TAO
1
;
Deng-Hao DONG
1
;
Min LANG
1
;
Shuang CHEN
1
;
Ying FENG
1
;
Min WU
1
Author Information
1. Department of Ophthalmology, Army Medical Center of PLA(Daping Hospital), Chongqing 400000, China
- Publication Type:Journal Article
- Keywords:
CASIA2;
IOL Master 700;
cataract;
corneal refractive power;
astigmatism
- From:
International Eye Science
2023;23(6):1017-1023
- CountryChina
- Language:Chinese
-
Abstract:
AIM: To compare the difference and consistency of corneal refractive power and astigmatism measured by CASIA2 and IOL Master 700 in patients with age-related cataract.METHODS: Retrospective study. A total of 153 patients(232 eyes)with age-related cataract admitted to Daping hospital from November to December 2021 were selected. The flat keratometry(Kf), steep keratometry(Ks), mean keratometry(Km), degree and axis of astigmatism(vector representation J0 and J45)of the anterior, posterior surfaces together with the total cornea from cataract patients were measured by CASIA2 and IOL Master 700, respectively. The difference, correlation and consistency of the two instruments were analyzed.RESULTS:There was no significant difference in J45 values of posterior corneal surface measured by CASIA2 and IOL Master 700(-0.006±0.038D vs. -0.005±0.044D, P>0.05), but there were significant differences in other parameters(all P<0.05). All parameters measured by the two instruments were significantly positive correlated(all r/rs>0.7, P<0.001); Bland-Altman analysis showed that the refractive power and astigmatism of the anterior cornea surface measured by the two facilities were in good consistency, while the refractive power of the posterior surface and the whole cornea showed poor consistency.CONCLUSION: CASIA2 and IOL Master 700 showed little differences and good consistency in the refractive power and astigmatism of the anterior, posterior and total corneal surface in cataract patients, which seems interchangeable. However, the refractive power of the posterior surface and the whole cornea has significant differences and poor consistency, which should not be interchange casually.