Evaluation of PET Mainstream Scattering Correction Methods.
10.3969/j.issn.1671-7104.2023.01.008
- Author:
Zhipeng SUN
1
;
Ming LI
1
;
Jian MA
1
;
Jinjin MA
1
;
Guodong LIANG
1
Author Information
1. Neusoft Medical Systems Co. Ltd., Shenyang, 110167.
- Publication Type:Journal Article
- Keywords:
Monte Carlo algorithm;
PET;
image reconstruction;
scattering correction
- MeSH:
Positron Emission Tomography Computed Tomography;
Scattering, Radiation;
Computer Simulation;
Brain;
Monte Carlo Method;
Phantoms, Imaging;
Image Processing, Computer-Assisted
- From:
Chinese Journal of Medical Instrumentation
2023;47(1):47-53
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVE:Current mainstream PET scattering correction methods are introduced and evaluated horizontally, and finally, the existing problems and development direction of scattering correction are discussed.
METHODS:Based on NeuWise Pro PET/CT products of Neusoft Medical System Co. Ltd. , the simulation experiment is carried out to evaluate the influence of radionuclide distribution out of FOV (field of view) on the scattering estimation accuracy of each method.
RESULTS:The scattering events produced by radionuclide out of FOV have an obvious impact on the spatial distribution of scattering, which should be considered in the model. The scattering estimation accuracy of Monte Carlo method is higher than single scatter simulation (SSS).
CONCLUSIONS:Clinically, if the activity of the adjacent parts out of the FOV is high, such as brain, liver, kidney and bladder, it is likely to lead to the deviation of scattering estimation. Considering the Monte Carlo scattering estimation of the distribution of radionuclide out of FOV, it's helpful to improve the accuracy of scattering distribution estimation.