Comparison of Ablation Performance between Octopus Multipurpose Electrode and Conventional Octopus Electrode
- Author:
Sae-Jin PARK
1
;
Jae Hyun KIM
;
Jeong Hee YOON
;
Jeong Min LEE
Author Information
- Publication Type:Original Article
- From:Korean Journal of Radiology 2023;24(2):86-94
- CountryRepublic of Korea
- Language:English
-
Abstract:
Objective:To compare Octopus multipurpose (MP) electrodes, which are capable of saline instillation and direct tissue temperature measurement, and conventional electrodes for radiofrequency ablation (RFA) in porcine livers in vivo.
Materials and Methods:Sixteen pigs were used in this study. In the first experiment, RFA was performed in the liver for 6 minutes using Octopus MP electrodes (n = 15 ablation zones) and conventional electrodes (n = 12 ablation zones) to investigate the effect of saline instillation. The ablation energy, electrical impedance, and ablation volume of the two electrodes were compared. In the second experiment, RFA was performed near the gallbladder (GB) and colon using Octopus MP electrodes (n = 12 ablation zones for each) with direct tissue temperature monitoring and conventional electrodes (n = 11 ablation zones for each). RFA was discontinued when the temperature increased to > 60°C in the Octopus MP electrode group, whereas RFA was performed for a total of 6 minutes in the conventional electrode group. Thermal injury was assessed and compared between the two groups by pathological examination.
Results:In the first experiment, the ablation volume and total energy delivered in the Octopus MP electrode group were significantly larger than those in the conventional electrode group (15.7 ± 4.26 cm3 vs. 12.5 ± 2.14 cm3 , p = 0.027; 5.48 ± 0.49 Kcal vs. 5.04 ± 0.49 Kcal, p = 0.029). In the second experiment, thermal injury to the GB and colon was less frequently noted in the Octopus MP electrode group than that in the conventional electrode group (16.7% [2/12] vs. 90.9% [10/11] for GB and 8.3% [1/12] vs. 90.9% [10/11] for colon, p < 0.001 for all). The total energy delivered around the GB (2.65 ± 1.07 Kcal vs. 5.04 ± 0.66 Kcal) and colon (2.58 ± 0.57 Kcal vs. 5.17 ± 0.90 Kcal) were significantly lower in the Octopus MP electrode group than that in the conventional electrode group (p < 0.001 for all).
Conclusion:RFA using the Octopus MP electrodes induced a larger ablation volume and resulted in less thermal injury to the adjacent organs compared with conventional electrodes.