An investigation of radiation dose to adult subjects in computed tomography scanning in Shanghai, China
10.13491/j.issn.1004-714X.2022.05.012
- VernacularTitle:上海地区CT扫描检查成年人受检者剂量分析
- Author:
Aijun QIAN
1
;
Linfeng GAO
1
;
Jie YAO
1
;
Xuexin WEN
2
;
Bin WANG
1
;
Hong XIAO
1
;
Zhou JIANG
1
Author Information
1. Shanghai Municipal Center for Disease Control and Prevention, Shanghai 200336 China.
2. Hongkou District Center for Disease Control and Prevention, Shanghai 200082 China.
- Publication Type:Journal Article
- Keywords:
CT;
Radiation dose;
Diagnostic reference level;
Dose length product
- From:
Chinese Journal of Radiological Health
2022;31(5):587-591
- CountryChina
- Language:Chinese
-
Abstract:
Objective To investigate dose parameters in computed tomography (CT) scanning of common body parts inall public medical institutions in Shanghai, China, and analyze the dose distribution of CT scanning among adult subjects in Shanghai, and to provide a basis for establishing the diagnostic reference levels of CT scanning in Shanghai. Methods We selected at least one medical institution with CT services each from all 16 districts of Shanghai. In each medical institution, a piece of CT equipment with qualified annual inspection was sampled to investigate the doses to subjects in head, chest, abdomen, and lumbar spine scanning. We collected the basic information of the subjects, CT scanning parameters, volumetric CT dose index (CTDIvol), and dose length product (DLP). Results The scanning data of 1475 subjects were obtained. The 75th percentiles of the CTDIvol of the head, chest, abdomen, and lumbar spine were 57 mGy, 11 mGy, 16 mGy, and 23 mGy, respectively. The 75th percentiles of the DLP were 862 mGy·cm, 361 mGy·cm, 593 mGy·cm, and 550 mGy·cm, respectively. Conclusion CTDIvol and DLP differed significantly at different body parts, and also differed greatly at the same body parts. The DLP of men was slightly higher than that of women.