Absolute measurement of surface particle emission rates from α and β plane sources using a digitizer
10.13491/j.issn.1004-714X.2023.01.001
- VernacularTitle:利用数字化仪实现 α、β 平面源表面粒子发射率的绝对测量
- Author:
Tianxiang LU
1
;
Qianqian ZHOU
2
;
Ming ZHANG
2
;
Fei TUO
1
;
Shuying KONG
1
Author Information
1. National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention
2. National Institute of Metrology
- Publication Type:Journal Article
- Keywords:
Digitizer;
Metrology;
Proportional counter;
Plane source;
Emission rate
- From:
Chinese Journal of Radiological Health
2023;32(1):1-5
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To prove the validity and accuracy of the digitizer instead of the conventional electronics plug-in
for radionuclide measurement.
Methods: Based on a large-area flow-gas multi-wire proportional counter for 2πα and 2πβ surface particle emission rate measurement, the DT5730 digital waveform sampler developed by CAEN was used for waveform signal acquisition, amplitude analysis, and data processing of the α-plane source 241Am and the β-plane source nuclides 14C, 36Cl, and 90Sr-90Y of different energies.
Results:The deviations between the α and β surface particle emission rate results obtained after dead time and background corrections and the measurements obtained based on the plug-in calibrator were
all within 0.6%, within the uncertainty range, under consistent experimental conditions such as electronics threshold and high pressure.
Conclusion:The digitizer is an effective alternative to conventional electronics plug-ins for α and β signal acquisition and processing and the accurate measurement of α and β emission rates.
- Full text:利用数字化仪实现α、β平面源表面粒子发射率的绝对测量.pdf