Reliability of OneCeph Cephalometric Analysis Application on the Devices with Different Screen Size
10.21315/aos2022.1701.OA09
- Author:
Pimsiri Kanpittaya
1
Author Information
1. Department of Orthodontics, Faculty of Dentistry, Chulalongkorn University, Bangkok 10330, Thailand
- Publication Type:Journal Article
- Keywords:
Application;
Cephalometric analysis;
OneCeph;
Reliability;
Screen size
- MeSH:
Cephalometry--instrumentation
- From:Archives of Orofacial Sciences
2022;17(1):137-150
- CountryMalaysia
- Language:English
-
Abstract:
ABSTRACT:This study aimed to assess the reliability of the OneCeph application according to personal
computer (PC), tablet and smartphone screen size in comparison with Dolphin software on
PC as a gold standard. Cephalometric landmarks were identified on 100 digital radiographs.
Twenty-four cephalometric measurements were made with Dolphin software as a gold standard
comparing with OneCeph application on smartphone (OS), OneCeph on PC (OP) and
OneCeph on tablet (OT). All measurements were repeated after four weeks for intra-examiner
reliability with intraclass correlation coefficients (ICC). One-way ANOVA and Kruskal Wallis
test were done for measurement comparison between methods (Dolphin, OS, OP and OT).
Results for OneCeph on smartphone and tablet, 21 measurements were comparable with
Dolphin while other three (NLA, H-angle and UL to S-plane) were not. In OneCeph on PC,
20 measurements were comparable with Dolphin while other four (NLA, H-angle, U1 to A-point
and UL to S-plane) were not. All different measurements were clinically insignificant except H-angle.
Intra-examiner reliability represented ICC above 0.9. In conclusion, OneCeph application on three
different screen size is reliable to use for cephalometric measurement. Most of the measurements are
comparable with gold standard and adequate to be utilised in clinical routine. OneCeph on smartphone
and tablet are advantageous from the portable feature over PC.
- Full text:2.2022my0012.pdf