Inter-rater Reliability of Myoton-3 Myometer for Assessing Muscle Tone in Healthy Adults
10.3969/j.issn.1006-9771.2013.11.017
- VernacularTitle:Myoton-3 肌肉检测仪在健康成人肌张力测量中的评价者间信度
- Author:
Hongmei WEN
;
Yue LAN
;
Zulin DOU
;
Lichen CHEN
;
Wenxia HONG
- Publication Type:Journal Article
- Keywords:
Myoton-3, muscle tone, inter-rater reliablity
- From:
Chinese Journal of Rehabilitation Theory and Practice
2013;19(11):1058-1060
- CountryChina
- Language:Chinese
-
Abstract:
Objective To examine the inter- rater reliablity of Myoton-3 Myometer in assessment of muscle tone in healthy subjects.Methods 20 healthy volunteers were assessed their muscle tone of biceps brachii and flexor carpi radialis muscles in relaxed state with Myoton-3 by 2 testers within 24 h. The frequency of damping oscillations (F value) measured by Myoton-3 was as the characteristic of the muscletone. The intraclass correlation coefficient (ICC) and Bland-Altman analysis were performed. Results The ICC was 0.72~0.88 and 0.79~0.89 in triple scan and ten-time scan pattern, respectively. The Bland-Altman analysis revealed no systematic errors between testers. ConclusionThe Myoton-3 Myometer is reliable between testers for measuring the muscle tone in healthy adults.