Measure Static Balance Function with Active Balancer EAB-100: Reference Values and Clinical Significance
10.3969/j.issn.1006-9771.2015.09.018
- VernacularTitle:平衡仪静态平衡功能参数正常参考值的建立及权重分析
- Author:
Huiyun SONG
;
Pu ZHANG
;
Xiaoping YUN
- Publication Type:Journal Article
- Keywords:
static balance, assessment, reference value, principal component analysis. Active Balancer EAB-100
- From:
Chinese Journal of Rehabilitation Theory and Practice
2015;21(9):1069-1073
- CountryChina
- Language:Chinese
-
Abstract:
Objective To establish the reference values of static balance function with Active Balancer EAB-100 in healthy Chinese adults, and to screen the valuable indices for clinical work. Methods There were 227 healthy adults measured with Active Balancer EAB-100 when standing still with eyes opened and closed. All the data recorded were analyzed with SPSS 19.0, and the weight of every parameter was obtained with principal component analysis. Results The reference values of parameters for static balance function were established in healthy Chinese adults. The loadings of whole path length, rectangle area, circumference area, effective value area, deflection center displacement X, deflection center displacement Y, rectangle area of Romberg rate and circumference area of Romberg rate were more than average. Conclusion It is need to focus on whole path length, circumference area, deflection center displacement X, deflection center displacement Y, and effective value area and circumference area of Romberg rate much more in clinical practice for the static balance assessment.