Design and function evaluation of 3D-printed hemi-elbow-arthroplastic total humeral prosthesis
10.3760/cma.j.cn115355-20210206-00080
- VernacularTitle:3D打印半肘关节型全肱骨假体的设计及功能评估
- Author:
Liang YAN
1
;
Sen DONG
;
Haijie LIANG
;
Xiaodong TANG
;
Taiqiang YAN
;
Jie ZANG
;
Zhiye DU
;
Wei GUO
;
Yi YANG
Author Information
1. 北京大学人民医院骨与软组织肿瘤治疗中心,北京 100044
- Keywords:
Bone neoplasms;
3D printing;
Total humerus;
Elbow joint;
Bionic hemiarthroplasty
- From:
Cancer Research and Clinic
2022;34(5):346-351
- CountryChina
- Language:Chinese
-
Abstract:
Obiective:To investigate the design, the key points of surgery and the outcome of short-term follow-up of hemi-elbow-arthroplastic bionic total humeral prosthesis by using 3D-print.Methods:The clinical data of 8 patients with humerus malignant tumor who underwent total humerus resection and a 3D-printed auxiliary hemi-elbow-arthroplastic total humeral prosthesis replacement from March 2018 to March 2020 at Peking University People's Hospital were retrospectively analyzed. The operative data, oncological outcome, complication profiles of prosthesis and functional status of all 8 patients were also analyzed. There were 3 males and 5 females with a mean age of 5-37 years. Histological diagnosis included 5 cases of osteosarcoma, 2 cases of Ewing's sarcoma and 1 case of chondrosarcoma.Results:The time for production of the prosthesis was (9.8±2.7) d, the operation time was (209±23) min and intraoperative hemorrhage was (569±173) ml. All the prostheses were implanted successfully and no patients experienced intraoperative complications. While 2 patients had postoperative complications, 1 case of temporary palsy of radial nerve and 1 case of local recurrence. LARS artificial ligament or hernia patch was used to reconstruct joint capsule and tendon-ligament attached around the elbow and shoulder joint. The flexion and extension of the elbow was (118±15)° (100-140°) and (11±9)° (0-25°), and the abduction and anteflexion of the shoulder was (28±12)° (15-50°) and (26±9)° (15-40°), respectively. The postoperative Musculoskeletal Tumor Society (MSTS)-93 scale score was (24.1±1.5) scores. The median follow-up time was 17 months (12-32 months), 7 patients had disease-free survival and 1 patient survived with tumor.Conclusions:The novel 3D-printed total humeral prosthesis with hemi-elbow-arthroplasty has a good perioperative safety, which is effective in restoring the function of elbow joint, solving the problem of stress concentration of ulnar marrow lever of total elbow joint prosthesis and lowering long-term wear rate and loosening rate of prosthesis.