Establishment and performance evaluation of Der p 1/Der p 2 specific immunoglobulin E detection for the components of dust mite allergens based on chemiluminescence immunoassay
10.3760/cma.j.cn321828-20200817-00314
- VernacularTitle:尘螨过敏原Der p 1、Der p 2特异性IgE化学发光免疫检测方法的建立及其性能评估
- Author:
Feifei HAN
1
;
Ying ZHOU
;
Jian ZHANG
;
Hao YIN
;
Meili WU
;
Xi CHEN
;
Yubao CUI
Author Information
1. 南京医科大学附属无锡人民医院医学检验科,无锡 214023
- Keywords:
Immunoglobulin E;
Antigens, dermatophagoides;
Dermatophagoides pteronyssinus;
Chemiluminescent measurements;
Nanoparticles
- From:
Chinese Journal of Nuclear Medicine and Molecular Imaging
2022;42(3):160-164
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To establish a quantitative detection method for the main components of dust mite allergens Der p 1, Der p 2 specific immunoglobulin E (sIgE) by using the nano-magnetic particle chemiluminescence immunoassay.Methods:The performance indexes of the established method were evaluated after setting up and optimizing the chemiluminescence detection system and immune reaction conditions of sIgE for dust mite allergen. Serum sIgE levels of 50 suspected allergic patients with dust mite were determined by this chemiluminescence method. At the same time, this method was compared with the Phadia kit and the consistency was analyzed by Kappa test. Results:The optimal amount of magnetic beads was 25 μg, the optimal reaction buffer (pH=7.4) contained 0.1 mol/L Tris-HCl and 0.25%( W/ W) casein, the optimal coating solution contatined 20 mmol/L phosphate buffer (PB) and 1%( W/ W) bovine serum albumin (BSA), and the luminescence enhancement solution contained 0.05%( V/ V) Triton X-100. The two-step immunoreaction was adopted, and the detection could be completed with 20 μl sample at the optimal reaction temperature of 37℃. The limit of detection (LOD) of the established nano-magnetic particle chemiluminescence system in detecting Der p 1 and Der p 2 sIgE antibodies were both less than 0.01 kU/L, with the linear range of 0.2-100.0 kU/L, the precision of less than 7%, and the cross contamination rate of 0.19% and 0.21%. Compared with the Phadia system, the positive and negative coincidence rate of Der p 1 were 78.0%(32/41) and 9/9 with good consistency ( Kappa=0.65, P=0.008), and the positive and negative coincidence rate of Der P 2 were 93.3%(28/30) and 85.0%(17/20) with good consistency ( Kappa=0.79, P=0.003). Conclusion:The nano-magnetic particle chemiluminescence immunoassay is successfully established for detecting dust mite allergen sIgE, which has good detection performance and good consistency with Phadia system.