The experimental study of the effect of detector combination on image resolution in wide-detector CT
10.3760/cma.j.cn112271-20220217-00058
- VernacularTitle:宽体探测器CT探测器组合对影像分辨力影响的实验研究
- Author:
Senlin GUO
1
;
Yue REN
;
Yongxian ZHANG
;
Zongrui ZHANG
;
Binbin YU
;
Dandan LIU
;
Yantao NIU
Author Information
1. 首都医科大学附属北京同仁医院放射科,北京 100730
- Keywords:
Wide-detector CT;
Contrast resolution;
Scanning mode
- From:
Chinese Journal of Radiological Medicine and Protection
2022;42(5):396-399
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To investigate the effect of different scanning modes, detector width and location in detector on high and low contrast resolution of wide-detector CT image.Methods:The Catphan600 phantom with high and low contrast resolution modules was scanned with GE Revolution CT at the same CTDI vol. The scans were performed with the detector widths of 40, 80 and 160 mm for sequential scanning mode and with the detector width/pitch combinations of 40 mm/0.516, 40 mm/0.984, 80 mm/0.508 and 80 mm/0.992 for spiral scanning mode. The resolution modules were placed at the adjacent region between two sequential scans, central and foot side edge in the longitudinal scanning range seperately. The subjective evaluation of the high and low contrast resolution was performed by two radiologists. Results:The high contrast resolution was 8 LP/cm at adjacent region between two sequential scans with the detector width of 80 mm or 160 mm in sequential scanning mode, and at the pitch of 0.5 in spiral scanning mode, while it was 7 LP/cm for the rest of detector combinations. The distinguishable diameter was 3 mm at 1% low contrast resolution at foot side edge with the detector widths of 80 mm or 160 mm in the sequential scanning mode, and it was 2 mm for all the other conditions. The distinguishable diameter was 2 mm at 1% low contrast resolution with the detector width of 40 mm and pitch 0.516 in the spiral scanning mode and it was worse with the wider detector and larger pitch.Conclusions:For the wide-detector CT, scanning mode, detector width, location in detector and pitches will affect the high and low contrast resolution to some degree. Appropriate selection should be done according to actual needs in clinical practice.