Medium-term follow-up of percutaneous pulmonary valve implantation using the Venus-P self-expanding valve
10.3760/cma.j.cn101070-20200902-01453
- VernacularTitle:Venus-P自膨式肺动脉瓣经皮植入术的中期随访
- Author:
Shubo SONG
1
;
Taibing FAN
;
Yu HAN
;
Weijie LIANG
;
Bin LI
;
Kaiyuan WU
;
Lin LIU
;
Dong LIANG
;
Gejun ZHANG
;
Xiangbin PAN
Author Information
1. 郑州大学人民医院,郑州大学华中阜外医院儿童心脏中心,郑州 450000
- Keywords:
Percutaneous implantation;
Self-expanding pulmonary valve;
Pulmonary regurgitation;
Tetralogy of Fallot;
Follow-up
- From:
Chinese Journal of Applied Clinical Pediatrics
2022;37(2):98-102
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To evaluate the medium-term clinical and hemodynamic outcomes of percutaneous pulmonary valve implantation (PPVI) using a domestic Venus-P self-expanding valve in the treatment of severe pulmonary regurgitation after Tetralogy of Fallot (TOF).Methods:Retrospective study.From December 2017 to December 2020, 13 TOF patients with (17.8±4.7) years old and (50.2±12.3) kg underwent PPVI using the Venus-P self-expanding valve in the Department of Children′s Heart Center, Zhengzhou University People′s Hospital were recruited.The mean valve size was (31.9±3.1) mm.All patients received the transannular patch surgery and developed severe pulmonary regurgitation.After PPVI, 13 patients were followed up for at least 12 months.The operation-related complications, improvement of valve and heart function and the durability and reintervention of the Venus-P self-expanding valve were analyzed.The right ventricular end-diastolic volume index (RVEDVi) before and after operation was compared by the paired t-test, and the New York Heart Association (NYHA) class was compared by the paired Wilcoxon signed rank sum test. Results:PPVI was successfully performed in all 13 patients without death.At 6 months post-PPVI, cardiac magnetic resonance imaging findings showed that RVEDVi was significantly reduced[(145.7±9.6) mL/m 2vs.(100.2±12.2) mL/m 2, P<0.05], and the NYHA class was significantly improved ( P<0.05). One patient presented moderate pulmonary valve regurgitation at 12 months postoperatively.No vegetation was found on echocardiography, and blood culture was negative in this case.Six patients did not have postoperative pulmonary valve regurgitation, and the remaining presented mild or less pulmonary regurgitation.One patient had sudden ventricular tachycardia on the 6 th day postoperatively, which was converted to sinus rhythm after synchronous electrocardiography.Only one case underwent appendectomy 7 months after operation due to acute appendicitis, and the remaining did not require valve-related reintervention after implantation.During the follow-up for (22.8±8.0) months, no patients had perivalve leakage, stent migration and rupture.Complications like embolization and coronary artery compression were not reported. Conclusions:PPVI using the Venus-P self-expanding valve is safe and effective in patients with severe pulmonary valve regurgitation after TOF surgery, showing an acceptable medium-term follow-up outcome.Studies with a large sample size and long follow-up period are still needed to validate our findings.