- Publication Type:Journal Article
- Keywords: emphysema; fatigue performance; loop; lung volume reduction
- MeSH: Bronchoscopy/methods*; Emphysema/surgery*; Humans; Lung; Pneumonectomy/methods*; Pulmonary Emphysema/surgery*; Treatment Outcome
- From: Chinese Journal of Medical Instrumentation 2022;46(3):332-335
- CountryChina
- Language:Chinese
- Abstract: Lung volume reduction loop uses bronchoscopic lung volume reduction(BLVR) technology to compress and collapse the necrotic emphysema tissue and exhaust the internal gas to achieve the purpose of lung volume reduction to treat emphysema. After the lung volume reduction loop is implanted into the human body, the compressed part of the lung tissue tends to expand with breathing, which makes the lung volume reduction loop expand into a linear trend periodically. Fatigue resistance is one of the most important performance indexes of the lung volume reduction loop. In the paper, Z-direction vibration fatigue machine was used to simulate the changes of human respiratory cycle movement to test the fatigue performance of lung volume reduction loop, which can provide some reference for the test method of in vitro fatigue performance of lung volume reduction related products in the future.