Miniature Non-invasive Blood Pressure Measurement and Verification System.
10.3969/j.issn.1671-7104.2022.03.009
- Author:
Hang-Duo NIU
1
;
Si-Nian YUAN
1
;
Zi-Fu ZHU
1
;
Ji-Lun YE
1
;
Xu ZHANG
1
;
Hui YU
2
Author Information
1. Shenzhen Key Lab for Biomedical Engineering, Shenzhen, 518000.
2. Innovation Institute, Guangdong Biolight Medi-Tech. Co. Ltd., Zhuhai, 510320.
- Publication Type:Journal Article
- Keywords:
algorithm validation;
miniature;
oscillographic method;
sphygmomanometer
- MeSH:
Auscultation;
Blood Pressure/physiology*;
Blood Pressure Determination;
Oscillometry;
Sphygmomanometers
- From:
Chinese Journal of Medical Instrumentation
2022;46(3):278-282
- CountryChina
- Language:Chinese
-
Abstract:
Mercury sphygmomanometer based on traditional auscultation method is widely used in primary medical institutions in China, but a large amount of blood pressure data can not be directly recorded and applied in scientific research analysis, meanwhile auscultation data is the clinical standard to verify the accuracy of non-invasive electronic sphygmomanometer. Focusing on this, we designed a miniature non-invasive blood pressure measurement and verification system, which can assist doctors to record blood pressure data automatically during the process of auscultation. Through the data playback function,the software of this system can evaluate and verify the blood pressure algorithm of oscillographic method, and then continuously modify the algorithm to improve the measurement accuracy. This study introduces the hardware selection and software design process in detail. The test results show that the system meets the requirements of relevant standards and has a good application prospect.