Intra- and Inter-rater Reliability of Shear Wave Elastic Imaging Technique for Elastic Modulus Measurements of Muscle and Tendon
10.3969/j.issn.1006-9771.2018.10.015
- VernacularTitle:剪切波弹性成像技术量化评估肌腱肌肉弹性模量的信度
- Author:
Kun QIN
1
;
Yanan FENG
2
;
Yapeng LI
2
;
Suhong SHEN
2
;
Chunlong LIU
1
;
Zhijie ZHANG
2
Author Information
1. Clinical College of Acupuncture, Moxibustion and Rehabilitation, Guangzhou University of Chinese Medicine, Guangzhou, Guangdong 510000, China
2. Luoyang Orthopedic Hospital of Henan Province, Orthopedic Hospital of Henan Province, Luoyang, Henan 471000, China
- Publication Type:Journal Article
- Keywords:
shear wave ultrasound elastography, tendon, muscle, elastic modulus
- From:
Chinese Journal of Rehabilitation Theory and Practice
2018;24(10):1201-1205
- CountryChina
- Language:Chinese
-
Abstract:
Objective To explore the intra- and inter-rater reliability of shear wave elastography (SWE) for the patellar tendon and quadriceps femoris stiffness.Methods From October to November, 2017, the stiffness of the patellar tendon and quadriceps femoris of 20 healthy men was measured by SWE, and measured again five days later. The results were evaluated with interclass correlation coefficient (ICC).Results For patellar tendon, ICC=0.79 in intra-rater and inter-rater; for rectus femoris, ICC=0.71 in intra-rater and 0.73 in inter-rater; for vastus lateralis muscle, ICC=0.84 in intra-rater and 0.74 in inter-rater; for vastus medialis,ICC=0.95 in intra-rater and 0.94 in inter-rater.Conclusion It is a reliable and repeatable method to measure the stiffness of patellar tendon and quadriceps femoris by SWE.