In-vitro investigation of the mechanical friction properties of a computer-aided design and computer-aided manufacturing lingual bracket system under diverse tooth displacement condition
10.4041/kjod.2019.49.2.73
- Author:
Do Yoon KIM
1
;
Sang Woon HA
;
Il Sik CHO
;
Il Hyung YANG
;
Seung Hak BAEK
Author Information
1. Private Practice, Suncheon, Korea.
- Publication Type:Original Article
- From:The Korean Journal of Orthodontics
2019;49(2):73-80
- CountryRepublic of Korea
- Language:English
-
Abstract:
OBJECTIVE:The purpose of this study was to compare the static (SFF) and kinetic frictional forces (KFF) of a computer-aided design and computer-aided manufacturing lingual bracket (CAD/CAM-LB) with those of conventional LB (Con-LB) and Con-LB with narrow bracket width (Con-LB-NBW) under 3 tooth displacement conditions.
METHODS:The samples were divided into 9 groups according to combinations of 3 LB types (CAD/CAM-LB [Incognito], Con-LB [7th Generation, 7G], and Con-LB-NBW [STb]) with 3 displacement conditions (no displacement [control], maxillary right lateral incisor with 1-mm palatal displacement [MXLI-PD], and maxillary right canine with 1-mm gingival displacement [MXC-GD]; n = 6/group). While drawing a 0.016-inch copper or super-elastic nickel-titanium archwire with 0.5 mm/min for 5 minutes in a chamber maintained at 36.5℃, SFF and KFF were measured. The Kruskal-Wallis method with Bonferroni correction was performed.
RESULTS:The Incognito group demonstrated the highest SFF, followed by the 7G and STb groups ([STb-control, STb-MXLI-PD, Stb-MXC-GD] < [7G-MXC-GD, 7G-MXLI-PD, 7G-control] < [Incognito-MXLI-PD, Incognito-control, Incognito-MXC-GD]; p < 0.001). However, there were no significant differences in SFF among the 3 displacement conditions within each bracket group. Within each displacement condition, the Incognito group demonstrated the highest KFF, followed by the 7G and STb groups ([STb-control, STb-MXLI-PD] < Stb-MXC-GD < 7G-MXLI-PD < [7G-control, 7G-MXC-GD] < [7G-MXC-GD, Incognito-MXLI-PD, Incognito-control] < [Incognito-control, Incognito-MXC-GD]; p < 0.001). MXC-GD exhibited higher KFFs than MXLI-PD in the same bracket group.
CONCLUSIONS:The slot design and ligation method of the CAD/CAM-LB system should be modified to reduce SFF and KFF during the leveling/alignment stage.