Intraocular Pressure with the Mackay-Marg Electronic Applanation Tonometer in Normal Eyes.
- Author:
Jae Bong CHUN
1
;
Sang Min KIM
Author Information
1. Department of Ophthalmology, Kyung Hee University, School of Medicine, Seoul, Korea.
- Publication Type:Original Article
- MeSH:
Intraocular Pressure*;
Linear Models;
Reading
- From:Journal of the Korean Ophthalmological Society
1982;23(3):595-600
- CountryRepublic of Korea
- Language:Korean
-
Abstract:
The measurement of the intraocular pressure were made with the Mackay-Marg electronic applanation tonometer, as compared with standard Goldmann tonometer and Schiotz tonometer in Korean 114 normal eyes. The results were as follows: 1. The mean intraocular pressure of 114 normal eyes was 16.61 +/- 3. 77 mmHg with a Mackay-Marg tonometer. 2. There was significant differances between the Mackay-Marg tonometer and Goldmann tonometric values(p<0.005). The corelation coefficent(r) was 0.975, the linear regression was Y = 1.37 + 1.06X. 3. The standard deviation for Mackay-Marg tonometer was greater thandoldmann readings, and it was about 1.37 mmHg higher than Goldmann's. 4. There was significant differances between the Mackay-Marg tonometer and Schiotz tonometric values(p<0.005). The corelation coefficent(r) was 0.938, the linear regression was Y = 0.82X - 0.07.