In vitro biomechanical analysis of the second-generation dynamic anterior plate-screw system for quadrilateral area
10.3760/cma.j.cn121113-20200909-00552
- VernacularTitle:第二代动力化前路方形区钛板螺钉系统的体外生物力学研究
- Author:
Haiyang WU
1
;
Xianhua CAI
;
Qipeng SHAO
;
Ranran SHANG
;
Chengjing SONG
;
Ximing LIU
;
Guodong WANG
;
Yanjin LI
;
Ruibing FENG
;
Hongqi ZHANG
Author Information
1. 解放军中部战区总医院骨科,湖北省骨创伤救治临床医学研究中心,武汉 430070
- Keywords:
Acetabulum;
Fracture fixation, internal;
Bone screws;
Bone plates;
Mechanics
- From:
Chinese Journal of Orthopaedics
2021;41(21):1569-1578
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To introduce the standard screw implantation methods and to analyze the biomechanical stability of the second-generation dynamic anterior plate-screw system for quadrilateral area (DAPSQ).Methods:Six adult formalin-preserved corpses were selected to make a complete pelvic specimen. Further, the left high double-column fracture models were made and randomly fixed with second-generation DAPSQ or anterior reconstruction titanium plate and 1/3 tube buttress-plate (ARTPB). The specimens of intact pelvis (IP) group, DAPSQ group and ARTPB group were fixed on a Zwick Z100 material machine and loaded vertically with 200 N, 300 N, 400 N, 500 N, 600 N, 700 N, and 800 N in a simulated sitting position, respectively. The axial displacement and strain changes in the anterior and posterior columns were tested in the three groups. The stiffness was calculated accordingly.Results:The axial compression displacement in the three groups showed an increase trend as well with the vertical load increased from 200 N to 800 N ( F=68.581, P<0.001; F=91.795, P<0.001; F=33.819, P=0.002). The axial displacement in ARTPB group was significantly larger than that in DAPSQ group and IP group ( P<0.05), while the difference between DAPSQ and IP groups was not significant ( P>0.05). Under the vertical load of 600 N, the pelvic axial stiffness of IP group, DAPSQ group, and ARTPB group were 220.72±70.33 N/mm, 185.68±48.49 N/mm and 135.83±60.58 N/mm, respectively. The axial stiffness of ARTPB group was significantly lower than that in DAPSQ group and IP group ( t=5.345, P=0.003; t=6.443, P=0.001), while the difference between DAPSQ and IP groups was not significant ( t=2.138, P=0.086). There were no significant differences of the strain values in anterior column among the three groups during the load increasing from 200 N to 800 N ( P>0.05). With the load increasing from 500 N to 800 N, the strain values of the posterior column in ARTPB group were significantly greater than those of IP and DAPSQ groups ( P<0.05). However, the differences between IP and DAPSQ groups were not statistically significant in strain values of the posterior column ( P>0.05). Conclusion:Compared with anterior reconstruction titanium plate and 1/3 tube buttress-plate, acetabular double-column fracture model fixed with the second-generation DAPSQ has less axial compression displacement but with greater axial stiffness. The stress change in the posterior columns of the acetabulum is like in IP. Therefore, the second-generation DAPSQ has reliable biomechanical stability.