Application of independent dose verification of volumetric modulated arc therapy based on Monte Carlo
10.3760/cma.j.cn113030-20201010-00494
- VernacularTitle:基于蒙特卡罗实现容积调强弧形治疗计划独立剂量验证的应用研究
- Author:
Tiantian CUI
1
;
Bing LI
;
Ru LIU
;
Danhong DING
;
Dingjie LI
;
Zhaoyang LOU
;
Hong GE
Author Information
1. 郑州大学附属肿瘤医院/河南省肿瘤医院放疗科 450008
- Keywords:
Volumetric modulated arc therapy;
Monte Carlo;
Independent dose verification
- From:
Chinese Journal of Radiation Oncology
2021;30(12):1286-1291
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To develop a verification platform based on Monte Carlo (MC) for independent dose verification of volumetric modulated arc therapy (VMAT) plans.Methods:The head model including collimator of Varian TrueBeam linear accelerator was constructed by using EGSnrc/BEAMnrc, and the independent dose verification platform for the patients’ VMAT plans was built based on the head model and an in-house code. The percent depth dose (PDD) curves and off-axis ratios for different field sizes, the dose distribution of two irregular fields and three VMAT plans of the head and neck, chest, and pelvis were simulated using the platform. The simulated results of the PDD curves and the off-axis ratios of different field sizes were compared with the blue water measurement results. The difference between the irregular fields and the actual ArcCHECK measurements was also investigated. Besides, the differences among the MC simulated dose, TPS calculated dose and the ArcCHECK measured dose were analyzed by several methods, such as γ analysis and dose-volume histogram to verify whether the platform could be independently employed for dose verification.Results:The MC simulated results of PDD curves and off-axis ratios from 4 cm×4 cm to 40 cm×40 cm were in good agreement with the measured results. And the γ passing rates between the MC simulation and the ArcCHECK measurement for the irregular fields were above 98.1% and 99.1% for 3%/2 mm and 3%/3 mm, respectively. For VMAT plans of three patients, the γ results between the MC simulated dose and ArcCHECK measured dose were better than 93.8% and 95.9% under the criteria of 3%/2 mm and 3%/3 mm respectively. At the same time, the γ passing rates of nasopharyngeal, lung, and rectal cancers were 95.2%, 98.6% and 98.9% based on 3D γ analysis using TPS calculated dose and MC simulated dose under the criteria of 3%/3 mm; the passing rates of these three were 90.3%, 95.1% and 96.7% for 3%/2 mm, respectively.Conclusions:The simulation results of the MC-based verification platform developed in this study show a good agreement with the actual measurement results, and the simulation results are closer to the real dose distribution using the patients’ data. The preliminary results demonstrate that the platform can be used for accurate independent dose verification of VMAT plans.