The phantom study of bismuth shielding combined with organ tube current modulation in brain CT
10.3760/cma.j.issn.0254-5098.2021.05.012
- VernacularTitle:铋屏蔽联合器官管电流调制技术在颅脑CT检查中应用的体模研究
- Author:
Jingkun SUN
1
;
Gang PENG
;
Fajin LYU
;
Jie WANG
;
Rui JIN
;
Xie HE
Author Information
1. 重庆医科大学附属第一医院放射科 400016
- Keywords:
Computed tomography;
Bismuth shielding;
X-care;
Radiation dose
- From:
Chinese Journal of Radiological Medicine and Protection
2021;41(5):385-389
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To evaluate the application value of bismuth shielding combined with organ tube current modulation (X-care) in brain CT scanning by measuring the radiation dose of sensitive organs.Methods:The head and neck phantom was scanned with Siemens dual source CT at the same volume CT dose index (CTDI vol) by X-care, bismuth shielding and x-care combined with bismuth shielding, and by dual energy CT angiography (DE-CTA) with and without bismuth shielding. The CT values of cerebral vessels, adjacent brain tissues and cerebrospinal fluid and image noise were measured, and the contrast noise ratio of cerebral vessels and brain parenchyma was calculated. Organ dose equivalent ( HT) was calculated by placing thermoluminescent personal dosimeter (TLD), and CTDI vol and dose length product (DLP) were recorded after each scan. Results:Under the same CTDI vol, the mean values of HT, lens with X-care, Bi shielding and X-care combined with Bi shielding were(37.89 ± 2.00), (42.20 ± 2.96) and (28.21 ± 1.31) mSv, respectively, significantly lower than those of conventional sequence scanning( F=186.52, P<0.05). The values of HT, thyroid with Bi shielding and X-care combined with Bi shielding were (0.77 ± 0.07) and (0.89 ± 0.08) mSv, lower than those of routine brain scan and X-care( F=103.26, P<0.05). The values of HT, lens and HT, thyroidof DE-CTA with bismuth shielding were (11.56 ± 1.04) and (0.32 ± 0.03) mSv, respectively, significantly lower than those without bismuth shielding( t=5.07, P<0.05). There was no significant difference in noise and CNR in routine brain scan between with and without X-care, bismuth shielding and X-care combined with bismuth shielding. There was no significant difference in noise and CNR in dual energy CTA scanning between with and without Bi shielding. Conclusions:Using bismuth shielding and organ tube current modulation, we can significantly reduce organ dose of lens and thyroid during brain CT scanning without sacrificing the image quality.