Non-invasive ultrasound cardiac output monitor of hemodynamic parameters in neonates with different gestational age and birth weight
10.3760/cma.j.issn.2096-2932.2021.06.002
- VernacularTitle:不同胎龄及出生体重新生儿无创超声心输出量参数值测定
- Author:
Wenchao HONG
1
;
Xiaohui GONG
;
Gang QIU
;
Cheng CAI
;
Li MA
;
Chongbing YAN
Author Information
1. 上海市儿童医院,上海交通大学附属儿童医院新生儿科 200062
- Keywords:
Ultrasound cardiac output monitor;
Hemodynamic parameters;
Infant,newborn;
Birth weight;
Gestational age
- From:Chinese Journal of Neonatology
2021;36(6):8-11
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To study the correlations of neonatal hemodynamic parameters with gestational age (GA) and birth weight (BW) using non-invasive ultrasound cardiac output monitor (USCOM).Method:From March to September 2019, neonates with stable hemodynamics admitted to the Department of Neonatology of our hospital were enrolled in this prospective study. According to their GA, they were assigned into <29 w group, 29~33 w group, 34~36 w group and ≥37 w group. According to their BW, they were assigned into <1 000 g group, 1 000~1 499 g group, 1 500~2 499 g group and ≥2 500 g group. Cardiac output (CO), cardiac index (CI), stroke volume (SV), myocardial contractility (inotropy, INO), flow time corrected (FTC), systemic vascular resistance index (SVRI) and heart rate (HR) were measured using USCOM. The univariate linear regression method was used to analyze the correlation of hemodynamic parameters with different GA and BW.Result:A total of 120 neonates with stable hemodynamics were enrolled, including 69 males and 51 females. The average GA was (34.2±3.8)w and the average BW was (2 221±860) g. SV ( r=0.489, P<0.001), CO ( r=0.681, P<0.001), CI ( r=0.348, P<0.001), FTC ( r=0.266, P=0.003), INO ( r=0.446, P<0.001)and HR ( r=-0.322, P<0.001) showed significant linear correlations with GA. No linear correlation existed between SVRI ( r=-0.052, P=0.574) and GA. SV ( r=0.603, P<0.001), CO ( r=0.852, P<0.001), CI ( r=-0.390, P<0.001), INO ( r=0.576, P<0.001) and HR ( r=-0.440, P<0.001) showed significant linear correlations with BW. No significant linear correlations existed between SVRI ( r=-0.076, P=0.409) or FTC ( r=0.090, P=0.329) and BW. Conclusion:USCOM can monitor neonatal hemodynamic parameters in real-time.Hemodynamic parameters including SV, CO, CI and INO are significantly different among newborns with different GA and BW and these parameters are linearly correlated with GA and BW.